22 quick start menu – Liquid Controls HML210 IOM - LCMag User Manual
Page 22

HML 210
22
QUICK START MENU
The user has immediate access to the Quick Start menu when the converter is powered up
by pressing the key . If access to the quick start menu does not occur, then it has been
disabled through the function 8.6
The set up functions (batching, regulation, flow and SMART measure) regulate
automatically the values of filters for the operation set modifying optimally the parameters
related to the function chosen.
To enable one of
four types operation press the key while the function is highlighted,
to confirm the changes press the key .
Filters settings for different S.rate (sensor menu, not user modifiable):
Batching
Regulation
Flow measure
SMART measure
MFSLN
MFFLN
MFSLN
MFFLN
MFSLN
MFFLN
MFSLN
MFFLN
10Hz
0
0
10
0
20
0
100
10
20Hz
0
0
5
0
10
0
50
5
50Hz
0
0
2
0
4
0
20
2
80Hz
0
0
1.3
0
2.5
0
12.5
1.25
150Hz
0
0
0.7
0
1.3
0
6.7
0.67
See the functions description section for more details regarding filters.
See programming functions section
Automatic parameters optimization (see below)
Access to all functions