13. SINGLE & DUAL ANALOG OUTPUT OPTIONS
Two versions of an analog board may be installed in the meter at rear panel jack position J4,
adjacent to the signal conditioner board. Once installed, this board is recognized by the meter,
which will bring up the appropriate menu items for it. These will not be brought up if an analog
output board is not installed.
A single analog output version can be configured for unipolar 4-20 mA current, 0-20 mA
current or 0-10V voltage, or bipolar -10 to +10V voltage (with a 20V voltage swing). Unipolar or
bipolar operation is selected by a jumper.
A dual analog output version can be configured for two unipolar outputs, which can each be
4-20 mA, 0-20 mA or 0-10V. Current or voltage output is selected at each connector.
Unipolar current or voltage: Jumper a
Bipolar -10 to +10 voltage: Jumper b
No jumpers, only selections at the connectors.
With either board, current or voltage output is selected at the connector and in the Menu. The
low analog output (0 mA, 4 mA, 0V, or -10V) may be set to correspond to any low displayed
reading
_An_Lo
. The high analog output (20 mA, 0V or 10V) may be set to correspond to any
high displayed reading
_An_Hi
. The meter will then apply a straight line fit between these two
end points to provide an analog output scaled to the meter reading. The decimal point location
is fixed by the
dEC.Pt1
selection.
KEYSTROKES FOR SETUP OF SINGLE ANALOG OUTPUT BOARD
If the MENU
key does not work, see Section 9 “Enabling & Locking Out Menu Items.”
Press Menu
Key
Press Digit Select
Key
Press Value Select
Key
00
Calibration output
selection.
0
0-20 mA current output
1
0-10V voltage output
2
4-20 mA current output
3
-10V to+10V voltage output
An_SEt
Analog Output Setup.
Press
until AnSEt is
displayed.
03
Analog output source.
3
Must be set to Item #3.
_An_Lo
Low displayed value for
-10V, 0V, 0 mA, or 4 mA
0.00000 0.00000 0.00000
0.00000 0.00000 0.00000
Select digit to flash
Select
0
thru
9
for flashing digit.
_An_Hi
.
High displayed value for
10V or 20 mA output
0.00000 0.00000 0.00000
0.00000 0.00000 0.00000
Select digit to flash
Select
0
thru
9
for flashing digit.