Laurel Electronics LAUREATE SERIES 2 SERIAL INPUT METER & REMOTE DISPLAY User Manual
Page 19
19
Press
Menu
Press Digit
Select Key
Press Value Select
Key
00000
Relay state when alarm is
active.
0
Relay 3 on
Relay 4 on
1
Relay 3 off
Relay 4 on
2
Relay 3 on
Relay 4 off
3
Relay 3 off
Relay 4 off
00000
Alarm latching or non-
latching (auto reset).
(see Glossary)
0
Alarm 3 auto reset Alarm 4 auto reset
1
Alarm 3 latching
Alarm 4 auto reset
2
Alarm 3 auto reset Alarm 4 latching
3
Alarm 3 latching
Alarm 4 latching
00000
Alarm operates at and
above setpoint (active
high) or at and below
setpoint (active low).
(see Glossary)
0
AL3 active high
AL4 active high
1
AL3 active low
AL4 active high
2
AL3 disabled
AL4 active high
3
AL3 active high
AL4 active low
4
AL3 active low
AL4 active low
5
AL3 disabled
AL4 active low
6
AL3 active high
AL4 disabled
7
AL3 active low
AL4 disabled
8
AL3 disabled
AL4 disabled
00000
Hysteresis mode or band
deviation mode
(see Glossary)
0
AL3 band deviation AL4 band deviation
1
AL3 hysteresis
AL4 band deviation
2
AL3 band deviation AL4 hysteresis
3
AL3 hysteresis
AL4 hysteresis
ALS 34
Alarm Setup
for relays 3 & 4
if detected.
00000
Number of consecutive
readings in alarm zone to
cause an alarm.
0
One reading in the alarm zone will cause
an alarm. This entry is ignored by remote
display. Enter any number, such as 0.
Press
Menu
Press Digit
Select Key
Press Value Select
Key
dEUn1H
Alarm 1 hysteresis
DEUn2H
Alarm 2 hysteresis
DEUn1b
Alarm 1 band deviation
DEUn2b
Alarm 2 band deviation
dEUn3H
Alarm 3 hysteresis
DEUn4H
Alarm 4 hysteresis
DEUn3b
Alarm 3 band deviation
DEUn4b
Alarm 4 band deviation
0.00000 0.00000 0.00000
0.00000 0.00000 0.00000
Select digit to flash
Select
-9
thru
9
for flashing
first digit,
0
thru
9
for other
flashing digits. Alarms will
activate above the setpoint
by the value entered and
deactivate below the setpoint
by the value entered. See
Glossary.