Laurel Electronics LAUREATE SERIES 2 DIGITAL PANEL METER User Manual

Page 57

background image

57

Load Cell Input

Range

Resolution

Resistance

Zero Range

Span Range

Error

20.000 mV
50.000 mV
100.00 mV
250.00 mV
500.00 mV

1 µV

2.5 µV

5 µV

12.5 µV

25 µV

1 GΩ

-99999 to

99999

0 to ±99,999

0.01% of FS

±2 cts

Thermocouple Accuracy

Span Tempco .........................................................................................0.003% of reading/°C

...............................................................................

0.0015% of reading/°C for load cell meter

Zero Tempco ............................................................................................................ 0.2 µV/°C
Reference Junction Accuracy .....................................................................1°C, 10°C to 40°C

Dual & Quad Relay Options

Power to Relay Option ................................................................................Powered by meter
Setpoint Setup....................................................... Via front panel pushbuttons or RS232/485
Update Rate .................................................................................56/s at 60 Hz, 47/s at 50 Hz
Response to input signal (min) ..................................................................Display update rate
Input Signal (selectable)........................................................Filtered or unfiltered input signal
Actuation Modes (selectable) ...... Above or below setpoint, latching or non-latching, disabled
Output Time Delay (selectable) .................................................................... 1 to 128 readings
Front Panel Enable / Lockout Modes (selectable) ................1) Display and change setpoints

...................................................................................

2) Display but do not change setpoints

..................................................................................

3) Neither display nor change setpoints

Alarm Status Indication ..........................................................................2 or 4 red LED lamps
Status Indication Setup (selectable) .......................Lit when output is ON or OFF, or disabled

Form C, SPDT Relay Output:

AC Rating ......................................................................................................... 8A @ 240 Vac
DC Rating........................................................................................................... 8A @ 24 Vdc
Isolation rating between signal common and contacts ............................................... 250 Vac
Insulation dielectric strength between signal common and contacts ........................................

................................................................................... 3.5 kV ac for 5 sec, 2.3 kV ac for 1 min

Form A, SPST Solid State Relay Output:

AC Rating ..................................................................................................120 mA @ 140 Vac
DC Rating..................................................................................................120 mA @ 180 Vdc
Isolation rating between signal common and contacts ............................................... 250V ac
Insulation dielectric strength between signal common and contacts ........................................

................................................................................... 3.5 kV ac for 5 sec, 2.3 kV ac for 1 min

Analog Output Option

Power to Analog Output Option...................................................................Powered by meter
Output Levels ............................................................. 0-20 mA, 4-20 mA, 0-10V, -10 to +10V
Voltage Compliance, 0-20 mA Output .................................................. 12V (0-600 Ohm load)
Current Compliance, 0-10V, -10 to +10V Output.......................2 mA (5 kOhm or higher load)