beautypg.com

Laurel Electronics LAUREATE SERIES 2 DIGITAL PANEL METER User Manual

Page 37

background image

37

Press Menu
Select Key

Press Digit
Select Key

Press Value Select
Key

SEtuP

Meter Setup
(continued)

00_00

Control inputs 1 & 2

(continued)

8

1 = Function Reset 2 = Display Blank

9

1 = Hold

2 = Display Blank

A

1 = Peak or Valley 2 = Display Blank

B

1 = Tare

2 = Display Blank

C

1 = Valley Display 2 = Peak Display

D

1 = Tare

2 = Tare Reset

Both control inputs 1 & 2 set to 1 for selec-
tions

2

,

4

,

A

,

C

= Function Reset.

Both control inputs 1 & 2 set to 1 for selec-
tions

0

,

1

,

3

,

5

,

8

,

9

,

B

,

D

= Meter Reset.

0

0000

Operation as a

rate of change meter.
Extended meter only.

0

Not rate of change

3

Rate x 10

1

Rate x 0.1

5

Rate x 1000

2

Rate x 1

6

Rate x 10000

00000

Operation of front
panel PEAK button
and rear connector for
Peak or Valley Display

0

Peak Display. Also selects “Peak” in

“Peak or Valley” at connector above.

1

Valley Display. Also selects “Valley” in

“Peak or Valley” at connector above.

2

Peak (1st push), Valley (2nd push)

3

Front panel Tare

00000

Auto-tare selection

0

No auto-tare

1

Auto-tare

ConFG

Meter Configuration

00000

RTD or Ohms wiring

RTD

Ohms

00

3 or 4 wire

3 or 4 wire

01

Not allowed

Custom curve

10

2-wire read

2-wire read

11

2-wire short

2-wire short

FiLtr

Filtering

0

0000

Alarm filtering

0

Unfiltered output

1

Filtered output

00000

Peak & Valley filtering

0

Unfiltered Peak & Valley

1

Filtered Peak & Valley

00000

Display filtering

0

Display batch average every 16 readings

1

Display filtered signal

00000

Adaptive filter

threshold

0

Low adaptive filter threshold level

1

High adaptive filter threshold level