Basic Ethernet Board

No jumpers needed.

RS232 Board

e

- Normal operation.

f

- Slave display to RS232 from another meter.

g

- Pull-up resistor on RTS line.

Note:

Board is shipped with jumpers e and g installed

RS485 Board, Full Duplex Operation

b & d

- Installed on last meter in long cable run.

RS485 Board, Half Duplex Operation

a & c

- Installed for half duplex operation.

d

- Installed on last meter in line with long cable runs.

Note:

Board is shipped with no jumpers installed.

RS485-Modbus Board, Full Duplex Operation

b & e

- Bias jumpers should be installed on 1 board.

a & d

- Installed on last meter in long cable run.

RS485-Modbus Board, Half Duplex Operation

b & e

- bias jumpers installed on 1 board.

c & f

- installed for half duplex operation.

a

- installed on last meter in line with long cable runs.

Note:

Board is shipped with no jumpers installed.

Modbus

b

a

c

d

e

RJ45

RJ45

f

Ethernet-to-RS485 Converter Board
& USB-to-RS485 Converter Board

Full Duplex Operation

No jumpers for short cable runs.
Add b & d for long cable runs.

Half Duplex Operation

a & c

- Installed for half duplex operation.

d

- Installed on last meter in line with long cable runs.