Laurel Electronics LAUREATE SERIES 2 DIGITAL PANEL METER User Manual
Page 16
16
KEYSTROKES FOR SETUP
If the MENU
key does not work, see Section 9 “Enabling & Locking Out Menu Items.”
Press Menu
Select Key
Press Digit
Select Key
Press Value Select
Key
_
dC U
DC Volts
__
0.2U
__
2.0U
_
20.0U
200.0U
600.0U
0.2, 2, 20, 200, 660V FS
_
dC A
DC Amps
__
2.0a
_
20.0a
_
200.0a
_
5.0a
0.2, 20, 200 mA, 5A FS
_
InPut
Selection of signal
input type & range
_
rAtio
Strain gauge & ratio
__
0.2U
__
2.0U
_
20.0U
0.2, 2, 20V FS.
00_00
Display selection with
scale factor of 1.
0
4-1/2 digits (±20,000)
1
Remote display (±99,999)
2
4-1/2 digits, counts by 10 (±20,000)
3
3-1/2 digits (±2,000)
00 _00
Power line frequency
0
Noise minimized for 60 Hz
1
Noise minimized for 50 Hz
00_ 00
Scaling method
0
Scale and offset method
1
Coordinates of 2 points method
2
Reading coordinates of 2 points method
SEtuP
Meter Setup
00_00
Control inputs 1 & 2:
True = logic 1 (0V or
tied to digital ground)
False = logic 0 (5V or
open)
0
1 = Reset, 2 = Meter Hold
1
1 = Function Reset, 2 = Peak or Valley
2
1 = Hold, 2 = Peak or Valley Display
3
1 = Hold, 2 = Tare
4
1 = Peak or Valley Display, 2 = Tare
5
1 = Tare, 2 = Reset
6
1 = 1, 2 = 1, decimal point = XXXXX
1 = 0, 2 = 1, decimal point = XXXX
.
X
1 = 1, 2 = 0, decimal point = XXX
.
XX
1 = 0, 2 = 0, decimal point = XX
.
XXX
7
1 = 1, 2 = 1, decimal point = XXXX
.
X
1 = 0, 2 = 1, decimal point = XXX
.
XX
1 = 1, 2 = 0, decimal point = XX
.
XXX
1 = 0, 2 = 0, decimal point = X
.
XXXX
8
1 = Function Reset, 2 = Display Blank
9
1 = Hold, 2 = Display Blank
A
1 = Peak or Valley, 2 = Display Blank
B
1 = Tare, 2 = Display Blank
C
1 = Valley Display, 2 = Peak Display
D
1 = Tare, 2 = Reset Tare to Zero
Both inputs 1 and 2 set to logic 1 for selec-
tions
2
,
4
,
A
,
C
= Function Reset
Both inputs 1 and 2 set to logic 1 for selec-
tions
0
,
1
,
3
,
5
,
8
,
9
,
B
,
D
= Meter Reset