9.3.5 Global system view
This option allows you check various components of the instrument:
•
O2 calibration valve
•
Gas calibration valve
•
Reference volume valve
•
Main purge valve
•
Needle purge valve
•
On-off valves function
•
Regulating valves function
•
Ultrasound
•
Flow regulation
•
Manual displacement
•
Proportional valve
Check a component by pressing on the button/schematic (which will turn to green) and observe
the values being displayed.
Note:
If the gas flow rate is manually changed, be aware that the automatic purge procedure (every 5
minutes) will reset the gas flow rate back to 0.25 mL/s, which will have an impact on the values being
observed.
9.3.6 System initialization
Selecting this option initiates the standard system startup process, without having to power the
instrument off and on. See
for more information on system
startup.
9.3.7 End application
Select this option to perform an orderly shut down of the instrument. On completion, the
instrument can be powered OFF.
9.3.8 Service validation
This option lists the service due date for a number of options. These are user-defined (see
option
). As each service due date is passed, the Service done button will
be made available. Press this button after the service has been performed to reset the service
due date based on how the service has been configured.
9.3.9 Measurement head initialization
Use this option to return the measurement head to the home position by pressing the Initialize
motor
button.