Calibration
76
Note: If the entry of the zero point value takes longer than
6 seconds, the display will alternate between showing the
conductivity value and the temperature value.
5. Press
ENTER
to confirm.
6. The display will show the new cell constant value and the zero point
.
Press ENTER to confirm.
7. Remove and clean the sensor from the calibration solution.
8. Immerse the sensor in the process water.
9. The transmitter remains in HoLD mode. Press
ENTER
again.
The transmitter remains in HoLD mode for approximately
20 seconds after calibration and returns to the measure mode.
Note: Repeat the calibration when “Err” is displayed.
8.2 Temperature sensor adjustment
1. Press
CAL
, enter passcode: 1015,
ENTER
. CAL TMP will be
displayed for 3 seconds. The transmitter is ready for calibration.
2. Measure the process temperature with a thermometer.
3. Use the
UP ARROW
and
RIGHT ARROW
to enter the value of the
measured temperature from the external thermometer.
4. Press
ENTER
to save the value.
5. The transmitter remains in HoLD mode
.
Press
ENTER
again.
The transmitter remains in HoLD mode for approximately
20 seconds after calibration and returns to the measure mode.
Note: Repeat the calibration when “Err” is displayed.