Calibration
72
o1.HoLD menu option. The Sensoface icon is inactive. The calibration
mode indicator is displayed (
To exit the calibration mode at any time, press
CAL
and then
ENTER
.
The output current will be held for 20 seconds and the measured value
will be displayed.
8.1.1
Enter cell constant
1. Press
CAL
, enter passcode: 1100,
ENTER
. CAL CELL will be
displayed for 3 seconds.
2. Use the
UP ARROW
and
RIGHT ARROW
to enter the value of the
cell constant of the connected sensor. The conductivity value will be
displayed also.
Note: A cell constant change will change the conductivity value.
Note: If the entry of the cell constant takes longer than 6 seconds,
the display will alternate between showing the conductivity and
temperature value.
3. Press
ENTER
to save the value.
4. The transmitter displays the cell constant value and CELL. Press
ENTER.
5. The transmitter remains in HoLD mode. Press
ENTER
again.
The transmitter remains in HoLD mode for approximately
20 seconds after calibration and returns to the measure mode.
Note: Repeat the calibration when “Err” is displayed.