Operation—Foundation Fieldbus
51
6.2.2.3 Data processing
Use the L_TYPE parameter to apply a linearization function to the data.
•
Direct—data is sent directly from the TB to the AI without
processing. The units for the XD_SCALE and OUT_SCALE
parameters must be identical.
•
Indirect—data from the TB is linearly scaled from the input scale
(XD_SCALE) to the output scale (OUT_SCALE).
•
Indirect square root—data is rescaled from the input scale
(XD_SCALE) and recalculated using a root function. Then the
value is linearly scaled to the output scale (OUT_SCALE).
6.2.2.4 Alarms
The AI block can generate block alarms and limit alarms. Use the
ACK_OPTION parameter to specify if an alarm must be acknowledged.
When the measured value status is “bad”, the AI block BLOCK_ERR
parameter indicates an Input Failure.
•
Block alarms—a block error will be reported via the
BLOCK_ERR parameter (simulate active, input failure, block
configuration error, out of service (OOS)). The BLOCK_ALM
parameter sends the alarm status to the control system.
•
Limit alarms—the measured value OUT falls outside of the limit
values (HI_HI_LIM, HI_LIM, LO_LIM, LO_LO_LIM).
If an alarm occurs, evaluate the following bus parameters:
•
OUT parameter (currently measured value) in the Al block
•
LAST_ERROR parameter in the transducer block
•
SENSOFACE_STATUS parameter in the transducer block