Section 5 Operation—4–20 mA/HART
The following section describes how to operate the si792(x) E/T
transmitter.
5.1 Measure mode
The display shows the configured process variable (conductivity,
concentration, resistivity or salinity) and the temperature value in the
measuring mode.
•
To return to the measurement mode during calibration press
CAL
and
ENTER
.
•
To return to the measurement mode during configuration press
CONF
and
ENTER
.
Note: The waiting time for the stabilization of the measured-value is
approximately 20 seconds.
5.2 Configuration
Use the configuration mode to specify the sensor, range and other
parameters for the system as shown in
Table 6
.
1. Press CONF and enter 1200 to enter the configuration mode.
2. Use the arrow and enter keys to change the settings. All settings
and options are shown in
,
and
To exit the configuration mode at any time, press
CONF
and
ENTER
.
The output current will be held for 20 seconds and the measured value
will be displayed.
Table 6 Configuration menu
Code
Parameter
Passcode
o1
4–20 mA current output
1200
tc
Temperature compensation
AL.
Alarm settings