Section 8 calibration, 1 calibration – Hach-Lange SI792(x) E HACH GLI 3700 series SI792(x) T 7MA2200 and 8398 series User Manual User Manual
Page 73
71
Section 8
Calibration
8.1 Calibration
Important Note: Use a thermometer to make sure that the displayed
temperature is accurate.
Important Note: Enter the sensor data in the configuration menu
before calibration.
Important Note: The transmitter must be calibrated each time a new
sensor is connected.
Important Note: If measurements are taken in an area with a cross-
section < 110 mm, use a calibration beaker with the same size and
material (metal/plastic).
Note: Only qualified personnel should conduct the tasks described in
this section of the manual.
The transmitter is adjusted to the sensor through the calibration. Use
the following methods and passcodes for calibration as described in
Table 14
.
During calibration the transmitter remains in the Hold mode for safety
reasons. The loop current is frozen at the value specified in the
Table 14 Methods and passcodes for calibration
Method
Passcode
Display calibration information
0000
Enter cell constant
1100
Calibration with a known calibration solution
0110
1
1 Not available for Profibus PA and Foundation Fieldbus versions.
Calibration by product comparison/grab sample
1105
1
Zero calibration in air
1001
Zero calibration with calibration solution
Temperature sensor calibration
1015