beautypg.com

2 calibration with a calibration solution, Calibration – Hach-Lange SI792(x) E HACH GLI 3700 series SI792(x) T 7MA2200 and 8398 series User Manual User Manual

Page 75

background image

Calibration

73

8.1.2 Calibration with a calibration solution

Use a calibration solution with a known conductivity value to adjust the
transmitter to show the same value as the solution.

Note: Stabilize the temperature during calibration.

1. Press

CAL

, enter passcode: 0110,

ENTER

. CAL SOL will be

displayed for 3 seconds. The transmitter is ready for calibration.

2. Remove and clean the sensor.
3. Immerse the sensor in the calibration solution.
4. Use the

UP ARROW

and

RIGHT ARROW

to enter the known value of

the calibration solution.
Note: If the entry of the calibration solution value takes longer than
6 seconds, the display will alternate between showing the cell
constant and the temperature value.

5. Press

ENTER

to save the value.

6. The display will show the new cell constant. Press

ENTER

.

7. The transmitter remains in HoLD mode. Press

ENTER

again.

The transmitter remains in HoLD mode for approximately
20 seconds after calibration and returns to the measure mode.

The display will show the temperature-corrected value of the
calibration solution (unless temperature compensation is set to off
in the configuration menu).

Note: Repeat the calibration when “Err” is displayed.