Calibration
73
8.1.2 Calibration with a calibration solution
Use a calibration solution with a known conductivity value to adjust the
transmitter to show the same value as the solution.
Note: Stabilize the temperature during calibration.
1. Press
CAL
, enter passcode: 0110,
ENTER
. CAL SOL will be
displayed for 3 seconds. The transmitter is ready for calibration.
2. Remove and clean the sensor.
3. Immerse the sensor in the calibration solution.
4. Use the
UP ARROW
and
RIGHT ARROW
to enter the known value of
the calibration solution.
Note: If the entry of the calibration solution value takes longer than
6 seconds, the display will alternate between showing the cell
constant and the temperature value.
5. Press
ENTER
to save the value.
6. The display will show the new cell constant. Press
ENTER
.
7. The transmitter remains in HoLD mode. Press
ENTER
again.
The transmitter remains in HoLD mode for approximately
20 seconds after calibration and returns to the measure mode.
The display will show the temperature-corrected value of the
calibration solution (unless temperature compensation is set to off
in the configuration menu).
Note: Repeat the calibration when “Err” is displayed.