beautypg.com

6 cyclic measured value status, 7 measured value limits-limit bits, Operation—foundation fieldbus – Hach-Lange SI792(x) E HACH GLI 3700 series SI792(x) T 7MA2200 and 8398 series User Manual User Manual

Page 57

background image

Operation—Foundation Fieldbus

55

6.2.2.6 Cyclic measured value status

The cyclic measured value status is shown in

Table 10

.

6.2.2.7 Measured value limits—limit bits

The respective status bit is set when a condition occurs (

Table 11

). The

status bit is reset when the condition no longer exists.

Table 10 Cyclic measured value status

Priority Quality

Sub-status

Bin-coding
(no limit bits)

Hex-
coding

Low

Good

Good Non-Specific

10 00 00 00

0 x 80

Good Active Advisory Alarm

10 00 10 xx

0 x 88

Good Active Critical Alarm

10 00 11 xx

0 x 8C

Uncertain

Uncertain Non-Specific

01 00 00 xx

0 x 40

Last Usable Value (LUV)

01 00 01 xx

0 x 44

Substitute-Set

01 00 10 xx

0 x 48

Initial Value

01 00 11 xx

0 x 4C

Sensor Conversion Not Accurate 01 01 00 xx

0 x 50

Engineering Unit Violation

01 01 01 xx

0 x 54

Sub-Normal

01 01 10 xx

0 x 58

Bad

Non-Specific

00 00 00 xx

0 x 00

Sensor Failure

00 01 00 xx

0 x 10

Device Value

00 00 11 xx

0 x 0C

High

Out of Service

00 01 11 xx

0 x 1C

Table 11 Limit bit description

Bin coding of limit bits

Description

00

OK

01

Low-limited

10

High-limited

11

Constant