Installation
21
3.5
Electrical installation
DANGER
Explosion hazard. Do not connect any components that are not
specified for the device. Always defer to the Hazardous Location
Control Drawing.
Prerequisites:
•
Review the applicable control drawing or ATEX EC-Type
Examination certificate
•
Review the electrical code regulations
•
Review the regulations for electrical installations in hazardous
locations, if appropriate (e.g. EN 60079-10/EN60079-14;
94/9/EC directive; NEC; CEC; Profibus Technical Guidelines
2.091)
•
Remove power or confirm non-hazardous status before making
any connections
•
Confirm that the intrinsic safety of the device is maintained when
connected to other equipment such as a power supply unit.
3.5.1
Wire preparation
To remove the terminal blocks from the transmitter for sensor wiring:
1. Insert a flat screwdriver between the terminal block and the
transmitter body.
2. Use the screwdriver as a lever to lift the terminal block off the
connectors (see
Figure 3
).