Maintenance

6.3 Routine maintenance schedule

The maintenance schedule is given for standard applications.
Deviant applications may cause different maintenance intervals.

Table 6 Routine maintenance schedule

Description

3 months
(Customer

task)

6 months

(Service

task)

12 months

(Service

task)

24 months

(Service

task)

Visual check analytical compartment, manual cleaning if
necessary.

X

1

X

Check filter pads, clean/ replace if necessary, particularly on
fan side.

X

1

X

Check reagents, replace if necessary.

X

1

X

Check cleaning solution, replace if necessary.

X

1

X

Check maintenance counters.

X

1

X

Check standard solutions, replace if necessary.

X

1

X

Visual check of electrode, replace membrane cap and
electrolyte if necessary

X

1

X

Check system for air impermeability.

X

Function check of both fans.

X

Function check heating for analyser enclosure.

X

General function check.

X

Read out and analyse event log. Read out and check data log
if necessary.

X

Check electrode (slope with intact membrane cap: –55mV to
–67mV), check every 6 months after 12 months of use.

(X)

2

X

Replace pump head for air pump.

X

Check magnetic stir bar, replace if necessary.

X

Replace reagent pump.

X

Check cleaning pump and replace if necessary (check every
6 months after 12 months of use)

(X)

2

X

Check stirring motor, replace if necessary.

X

1

Recommended maintenance interval, especially for reagents. The actual reagent and electrolyte (AMTAX sc) exchange

intervals depend on configuration.

2

Maintenance cycles are given for standard applications. Deviant applications may cause different maintenance intervals.