2 sensor setup menu (continued), Operation – Hach-Lange AMTAX indoor sc User Manual
Page 39
39
Operation
CONFIGURE (continued)
SAMPLE DETECTION
OFF/WARNING/ERROR
Determines instruments reaction when amount of available sample is too low. When
instrument is in filter probe mode, deactivating the sample detection will switch a deactivated
"STATUS MODUL.ERR" to 14 %
EXHAUST CONTROL
ON/OFF
Determines instrument reaction when drain is blocked
ELECTROLYTE
WARNING/OFF
Determines whether a warning is initiated when electrolyte level is low/ last change of
membrane is more than 90 days ago.
FALSE ELEC DATA
ERROR/OFF
Determines whether an error is initiated when the mV-Zero value of the electrode is not within
a certain range (see list of errors)
DEFAULT SETUP
Resets factor and settings to factory settings.
LAST CHANGE
Indication of the last change of a setting in the configuration menu.
MAINTENANCE
INFORMATION
LOCATION 1
Indication of measuring location 1
LOCATION 2
Indication of measuring location 2 on two-channel version
TYPE
Indication of instrument type
SENSOR NAME
Indication of instrument name
SERIAL NUMBER
Indication of serial number
RANGE
Indication of measuring range
OPTION
Indication of instrument option (filtration probe/1-channel/2-channel)
SOFTWARE PROBE
Filtration probe software
SOFTW. AMTAX
Instrument software
LOADER
Detailed information on the software in the instrument
APPL
Detailed information on the software in the instrument
STRUCTURE
Detailed information on the software in the instrument
FIRMWARE
Detailed information on the software in the instrument
CONTENT
Detailed information on the software in the instrument
LANGUAGE
List of languages that are supported by the installed language package.
CALIB. DATA
LOCATION1
Indication of measuring location 1
GAIN CORR
Indicates the correction factor set for correcting the measured values at measuring location 1.
DATE
Indicates the date of the last change to the correction factor.
LOCATION2
Available with the 2-channel version.
GAIN CORR
Indicates the correction factor set for correcting the measured values at measuring location 2.
DATE
Indicates the date of the last change to the correction factor.
mV ZERO
Electrode signal at the zero point
mV STANDARD 1
Electrode signal with standard 1
mV STANDARD 2
Electrode signal with standard 2
mV SLOPE
Electrode signal change per decade
LAST CALIBRAT.
Time of the last calibration
mV ACTIVE
Current electrode potential
5.2 Sensor setup menu (continued)