beautypg.com

Fieldbus communications – Hach-Lange AMTAX indoor sc User Manual

Page 114

background image

114

Fieldbus Communications

DISCHARGE
CALIB.

40067

Unsigned

Integer

1

R/W

0/10

Discharged values after a
calibration

REMAINING TIME

40068

Unsigned

Integer

1

R

0/65535

Remaining time of the current
process

APPL.

40069

Float

2

R

0/3.4028

2347E+

38

Version of application file

HEATING ON

40071

Unsigned

Integer

1

R/W

0/1/2/3/4/
5/6/7/8/9/

10/11/12

Enter the month when the sample
line heating is switched ON;
0 = always OFF, 1=January,
2=February to 12=December

HEATING OFF

40072

Unsigned

Integer

1

R/W

1/2/3/4/5/

6/7/8/9/1

0/11/12

Enter month when the sample line
heating is switched OFF;
1=January, 2=February to
12=December

CLEANING
MODULES

40073

Time2

2

R

Date of last filter module cleaning

SET PARAMETER 1

40075

Unsigned

Integer

1

R/W

19/42

Select the parameter for channel
one; 19=NH4–N, 42=NH4

AIR FILTER DISPL

40076

Integer

1

R

-32768/3

2767

Due day for cleaning/exchange of
air filter pads.
Negative values show overdue
cleaning/exchange of air filter pads.

COOLING

40077

Unsigned

Integer

1

R

0/100

Percentage of cooling fan power

ANALYZER
HEATING

40078

Unsigned

Integer

1

R

0/100

Heating of the analyzer

INTERVAL

40080

Unsigned

Integer

1

R/W

0 to 23

Measurement intervall;
0=5 minutes, 1=10 minutes,
2=15 minutes to 23=120 minutes,
35=3h, 47=4h, 59=5h, 71=6h,
83=7h, 95=8h, 107=9h, 119=10h,
131=11h, 143=12h, 155=13h,
167=14h, 179=15h, 191=16h,
203=17h, 215=18h, 227=19h,
239=20h, 251=21h, 263=22h,
275=23h, 287=24h

CLEANING START

40081

Unsigned

Integer

1

R/W

0/1/2/3/4/
5/6/7/8/9/

10/11/12/

13/14/15/
16/17/18/
19/20/21/

22/23

Acceleration time for cleaning
(24 hour format) 0=0 o`clock to
23=23o´clock

STATUS MODULES

40082

Unsigned

Integer

1

R

0/100

Status of the modules as integer in
percent

NEW MODULES

40083

Time2

2

R/W

Date of the last filter module
exchange

CLEAN. INTERVAL

40085

Unsigned

Integer

1

R/W

0/1/3/6/8/

12/24

Cleaning intervall ; 0=OFF, 1=1h,
3=3h, 6=6h, 8=8h, 12=12h, 24=24h

Table 13 Sensor Modbus registers (continued)

Tag name

Register #

Data

type

Length

R/W

Discrete

range

Min/Max

range

Description

This manual is related to the following products: