Operation
5.2.1 System setup menu
For more information on System Setup (current outputs, relays and
network interfaces), refer to the sc1000 user manual.
5.3 Calibration process
Note: Make sure that all solutions are available to avoid incorrect
measurements.
1. To start an automatically calibration select
CALIBRATION>CALIBRATE>AUTOCAL>SET INTERVAL.
OR
1. To start a calibration manually select
CALIBRATION>CALIBRATE.
Note: Press START to confirm and start the calibration process.
After the next measurement and the 5 minutes waiting time, the
calibration will automatically start and proceed with all
necessary standards.
Depending on the number of standards requiring calibration, a
cycle may take up to 40 minutes. After a successful calibration, the
instrument automatically returns to the measurements.
Note: When the instrument detects and displays a warning, the
measurement process will continue. Refer to
troubleshoot the warning.
Note: When the instrument detects and displays an error, the instrument
stops measuring. Refer to
to troubleshoot the error.
5.4 Cleaning process
For information about the cleaning interval refer to
Note: Make sure that the cleaning solution is available that the instrument
can work properly.
1. To configure an automatical cleaning interval select
CONFIGURE>CLEANING>SET INTERVAL.
OR
1. To start a manual cleaning cycle select
MAINTENANCE>CLEANING.
Note: Press START to confirm and start the cleaning process.
A cleaning cycle may take up to 10 minutes and then the
instrument returns automatically to the measuring mode.