A.16 option 10a plumbing and connections, Plumbing and connection options – Hach-Lange AMTAX indoor sc User Manual
Page 102
102
Plumbing and Connection Options
A.16 Option 10a plumbing and connections
Option 10a uses an sc analyzer with any type of sample
preparation that delivers a continuous sample stream that cannot
be pressurized. The waste of the analyzer is discharged into an
open drain.
and the following instructions for Option 10a:
1. Install the sample preparation unit.
2. Feed the sample line from the sample preparation unit through
the analyzer (
, item 2). Use Seal Plug #3 to secure.
3. Feed the drain tube through the analyzer (item 3). Use Seal
Plug #3 to secure.
Note: Tubes can be pushed through prepared holes on Seal Plug #3.
4. Seal any unused holes with Seal Plug #3.
5. Connect the drain tube to the T-fitting.
6. Connect the sample line from the sample preparation to the
bottom inlet on the overflow vessel using the fittings.