3 fuse replacement, 2 reagent replacement, Maintenance – Hach-Lange AMTAX indoor sc User Manual

Page 47

background image

47

Maintenance

4. Remove the fan locking screw and slide the retaining strap to

the top and remove (

Figure 10, page 20

). If necessary, press

the fan down to remove the retaining strap.

5. Slide the fan from the holding screws.

6. Clean the filter with soap and water and reposition.

7. Press ENTER.

8. Replace the fan. Make sure that the fan opening is positioned

downward. Attach the retaining strap (hold down the fan) and
install the fan locking screw.

9. Close the analyzer enclosure and the analysis panel.

10. Press ENTER.

The instrument will reset the maintenance counter and will start the
analyzer again.

6.1.3 Fuse replacement

The fuses for the power supply are found in the sc1000 controller.
Refer to the sc1000 User Manual for fuse replacement information.

6.2 Reagent replacement

The chemicals must be changed or renewed at regular intervals.
Refer to

Table 5

for information on the life of the chemicals.

Table 5 Chemicals for the AMTAX sc

Chemical (

section 8.1,

page 65

)

Measuring range 1

(0.02–5 mg/L)

Measuring range 2

(0.05–20 mg/L)

Measuring range 3

(1–100 mg/L)

Measuring range 4

(10–1000 mg/L)

Reagent

2500 mL for 3 month

2500 mL for 3 months

2500 mL for 2 months

2500 mL for 2 months

Standards (2 liters):

0.5 and 2.5 mg/L for
2 month on daily
calibration

1 and 10 mg/L for
3 months on daily
calibration

10 and 50 mg/L for
3 months on daily
calibration

50 and 500 mg/L for
3 months on daily
calibration

Cleaning and calibration

250 mL for 1 month on 3 cleanings and calibrations per day (default)

250 mL for 3 months on daily cleaning and calibration

Electrolyte and
membrane cap

11 mL
Electrolyte for
1–1.5 months;
Membrane cap:
2–3 months
(depending on the
membrane soiling)

11 mL
Replace with
membrane every
2–3 months
(depending on the
membrane soiling)

11 mL
Replace with
membrane every
2–3 months
(depending on the
membrane soiling)

11 mL
Replace with
membrane every
2–3 months
(depending on the
membrane soiling)