9 connect the data network, Installation – Hach-Lange AMTAX indoor sc User Manual
Page 34
34
Installation
does not work with 24 V versions of the sc1000 because it does not
provide the appropriate connectors for the analyzers.
Refer to the sc1000 manual for more information on
connecting power.
Please note the input voltage of the device. The device is available
in two non-adjustable voltage variants (115 V or 230 V).
The output voltage supplied by the controller at the outlets
corresponds to the mains voltage that is customary in the country in
question and to which the controller is connected.
A device designed for 115 V must not be connected to a controller
with a higher mains voltage.
1. Remove the power socket from the sc controller.
2. Connect the plug from the AMTAX sc to the power socket on
the sc controller.
3.9 Connect the data network
Refer to the sc1000 controller user manual for more information on
connecting the data network (item 1,
Figure 19 Connect the AMTAX sc to the sc1000 power supply
1
Data connector
2
Power connectors