Fieldbus communications – Hach-Lange AMTAX indoor sc User Manual
Page 118
![background image](/manuals/332764/118/background.png)
118
Fieldbus Communications
EXHAUST
CONTROL
40272
Unsigned
Integer
1
R/W
0/1
—
Analyzer checks the exhaust
whether it is blocked; 0=OFF, 1=ON
ELECTROLYTE
40277
Integer
1
R
—
-32768/3
2767
Left days until next electrolyte
change, negative values shows
overdue
ELECTROLYTE
40278
Unsigned
Integer
1
R/W
0/1
—
Sets if there is a warning when
electrolyte has to be exchanged;
0=OFF, 1=WARNING
FALSE ELEC DATA
40279
Unsigned
Integer
1
R/W
0/1
—
Sets if there is an error if the ZERO
value is out of range; 0=OFF, 1=ON
ERROR LIST
40280
Unsigned
Integer
2
R
—
—
Errors are coded bit wise. bit
0=TEMP. < 0 °C/ 32°F?, bit
1=ANALYZ. TO COLD, COOLING
FAILED,HUMIDITY ANALY,
HUMIDITY PROBE; PROBE
MISSING, NO HEAT UP,
CUVSENSOR DEFECT,
TEMPSENS DEFECT, CUVHEAT
DEFECT, CUV TOO HOT,
ELECTRODE SLOPE, FALSE
ELEC DATA, MODULES CONTAM.,
DRAIN BLOCKED, SAMPLE1, bit
16= SAMPLE2
WARNING LIST
40282
Unsigned
Integer
2
R
—
—
Warnings are coded bit wise, bit 0=
WARMUP PHASE, bit 1=COOLING
DOWN, SERVICE MODE,
REAGENT LEVEL, CLEAN SOLU
LEVEL, ANALYZER TO COLD,
ANALYZER TO WARM, CUV TOO
COOL, MODULES CONTAM.,
STANDARDS LEVEL,
ELECTRODE SLOPE,
reserved,SAMPLE1,
ELEKTROLYTE, bit 14=SAMPLE2
EDIT NAME
40285
String
8
R/W
Name of LOCATION (see menu
system)
ELECTROLYTE
40293
Float
2
R
electrode drift per 24h in mV
Table 13 Sensor Modbus registers (continued)
Tag name
Register #
Data
type
Length
R/W
Discrete
range
Min/Max
range
Description