Fieldbus communications – Hach-Lange AMTAX indoor sc User Manual
Page 117

117
Fieldbus Communications
OPERATING
HOURS
40173
Unsigned
Integer
2
R
—
0/99999
999
Operating hours of analyzer
PUMP
MEMBR.DISP.
40177
Integer
1
R
—
-32768/3
2767
Remaining days for pump
membrane in filter probe
COMPRESSOR
40186
Integer
1
R
—
-32768/3
2767
Remaining days for air compressor
LAST
CHA.FACTOR1
40194
Time2
2
R/W
—
—
Date of the last correction factor for
channel 1
LAST
CHA.FACTOR2
40196
Time2
2
R/W
—
—
Date of the last correction factor for
channel 2
SAMPLE
DETECTION
40218
Unsigned
Integer
1
R/W
0/1/2
—
Output if sample detection detects
low sample amount; 0=Warning,
1=Error 2=OFF
ACTUAL
MEAS.TIME
40224
Time2
2
R
—
—
Time of actual measurement value
LAST TIME
40226
Time2
2
R
—
—
Time of last measurement value
2.ND LAST TIME
40228
Time2
2
R
—
—
2.ND LAST TIME
3.RD LAST TIME
40230
Time2
2
R
—
—
3.RD LAST TIME
4.TH LAST TIME
40232
Time2
2
R
—
—
4.TH LAST TIME
5.TH LAST TIME
40234
Time2
2
R
—
—
5.TH LAST TIME
6.TH LAST TIME
40236
Time2
2
R
—
—
6.TH LAST TIME
7.TH LAST TIME
40238
Time2
2
R
—
—
7.TH LAST TIME
8.TH LAST TIME
40240
Time2
2
R
—
—
8.TH LAST TIME
9.TH LAST TIME
40242
Time2
2
R
—
—
9.TH LAST TIME
ACTUAL VALUE
40244
Float
2
R
—
0/15000
Actual measurement value, not
channel dependent
LAST VALUE
40246
Float
2
R
—
0/15000 LIST OF VALUES
2.ND LAST VALUE
40248
Float
2
R
—
0/15000 LIST OF VALUES
3.RD LAST VALUE
40250
Float
2
R
—
0/15000 LIST OF VALUES
4.TH LAST VALUE
40252
Float
2
R
—
0/15000 LIST OF VALUES
5.TH LAST VALUE
40254
Float
2
R
—
0/15000 LIST OF VALUES
6.TH LAST VALUE
40256
Float
2
R
—
0/15000 LIST OF VALUES
7.TH LAST VALUE
40258
Float
2
R
—
0/15000 LIST OF VALUES
8.TH LAST VALUE
40260
Float
2
R
—
0/15000 LIST OF VALUES
9.TH LAST VALUE
40262
Float
2
R
—
0/15000 LIST OF VALUES
STAT.
MODUL.WAR.
40266
Unsigned
Integer
1
R/W
40/30/15
—
Configure the level of warning for
the status of the modules
STATUS
MODUL.ERR
40267
Unsigned
Integer
1
R/W
14/10/8/0
—
Configure the level of error for the
status of the modules
ENCLOSU.TEMP.
MAX
40268
Float
2
R
—
-50/200
Maximum temperature inside the
analyzer during the last 24 hours,
intervall starts with power on
ENCLOSU.TEMP.
MIN
40270
Float
2
R
—
-50/200
Minimum temperature inside the
analyzer during the last 24 hours,
intervall starts with power on
Table 13 Sensor Modbus registers (continued)
Tag name
Register #
Data
type
Length
R/W
Discrete
range
Min/Max
range
Description