Fluid Components International ST51 User Manual
Page 13
![background image](/manuals/330676/13/background.png)
ST51 MASS FLOW
FLUID COMPONENTS INTERNATIONAL LLC
This page is subject to proprietary rights statement on last page
13
Doc. No. 06EN003389 Rev. -
Calibration Parameters Verification
The instrument uses a set of predetermined calibration parameters to process flow signals. Most of these parameters should not
change. A data package located with this manual contains the “ST51 Delta R Data Sheet”. This contains the calibration parameters
stored in the flow transmitter at the factory. To verify that these parameters have not changed, complete the following:
1. Identify the appropriate Delta R Data sheets by serial number of the instrument.
2. Press [D] [ENTER] to examine each of the parameters. The [ENTER] key allows scrolling one message at a time. Use Table
4 to verify parameters with the Delta R Data sheet ST51 Parameters.
Table 4. Diagnostic Test Sequence on Display
S/W Vers ion:
dR M in:
T SpanIDA C 0:
Flow Factor:
dR M ax:
T ZeroIDA C 0:
Cmin Flow:
Cal Ref:
T SpanIDA C 1:
Cmax Flow:
Tcslp:
T ZeroIDA C 1:
Eng Units:
Tcslp 0:
State 0:
Line Size 0:
Tcslp 2:
Switch Pt 0:
Line Size 1:
Tot M enu:
State 1:
Cmin Temp:
Tot Flag:
Switch Pt 1:
Cmax Temp:
Totalizer:
K factor 1:
M in Flow:
Rollover Cnt:
K factor 2:
M ax Flow:
Fix Pt Flag:
K factor 3:
Dens ity:
Puls e Factor:
K factor 4:
*C1 [1]:
Puls e Out:
I factor:
*C1 [2]:
Hours:
Temp Flag:
*C1 [3]:
Sample Period:
Out M ode:
*C1 [4]:
dR Slope:
Boxcar Max:
*C1 [5]:
dr Off Set:
RTD-SLP-385:
Break Pt:
Refr Slope:
% of Range:
*C2 [1]:
Refr Off Set:
Us er Name:
*C2 [2]:
SpanIDA C 0:
Shop Order #:
*C2 [3]:
ZeroIDA C 0:
Serial No.:
*C2 [4]:
SpanIDA C 1:
M odel#:
*C2 [5]:
ZeroIDA C 1:
If parameters that have an asterisk (*) have changed, this may indicate a problem. Customer Service should be contacted. If the
parameters have not changed, continue with the next section.