INSTALLATION, OPERATION AND TROUBLESHOOTING GUIDE

Step 3a. Wiring Preparation (Cylindrical Enclosure Option)

A.

Loosen 1 Allen Screw and unscrew the bottom
cover. The customer termination board is
uncovered.

Before the instrument is opened to install the wiring, FCI recommends that the following ESD precautions be observed:

Use a wrist band or heel strap with a 1 megohm resistor connected to ground. If the instrument is in a shop setting there should
be static conductive mats on the work table and floor with a 1 megohm resistor connected to ground. Connect the instrument
to ground. Apply antistatic agents such as Static Free made by Chemtronics (or equivalent) to hand tools to be used on the
instrument. Keep high static producing items away from the instrument such as non-ESD approved plastic, tape and packing
foam.

The above precautions are minimum requirements to be used. The complete use of ESD precautions can be found in the U.S.
Department of Defense Handbook 263.

Open the instrument as shown to wire the instrument:

C.

For Remote Installations Only:
Loosen 1 Allen Screw and
unscrew the top cover. Access to
plug TS2 will be required to wire
the instrument per the remote
installation information on the
Page 7.

B.

Route conduit to the enclosure. Wire the
termination board per the wiring information
on the next pages.

D.

When finished wiring, replace the cover(s) and tighten
the Allen screw(s). See the Installation Section in the
ST98 Manual, Document Number 06EN003291 for more
details.

Step 3b. Wiring Preparation (Box Enclosure Option)

Before the instrument is opened to install the wiring, FCI recommends that the following ESD precautions be observed:

Use a wrist band or heel strap with a 1 megohm resistor connected to ground. If the instrument is in a shop setting there should
be static conductive mats on the work table and floor with a 1 megohm resistor connected to ground. Connect the instrument
to ground. Apply antistatic agents such as Static Free made by Chemtronics (or equivalent) to hand tools to be used on the
instrument. Keep high static producing items away from the instrument such as non-ESD approved plastic, tape and packing
foam.

The above precautions are minimum requirements to be used. The complete use of ESD precautions can be found in the U.S.
Department of Defense Handbook 263.

Open the instrument as shown on the next page to wire the instrument: