28

installation of the 8020 on the
s020 fitting

Insert nut 3 on fitting 5.

Insert snap ring 2 into
groove 4.

Check that seal 6 is
correctly inserted on the
flow sensor.

Slowly insert device 1
into the fitting.

If the mounting is correctly
done the device cannot be
turned around anymore.

Hand lock the assembly
with nut 3.

1

2

3

4

5

6

Fig. 2: Installation of the flowmeter on the S020 fitting

English

29

Wiring

danger

risk of injury due to electrical discharge
• Shut down and isolate the electrical power source

before carrying out work on the system.

• Observe all applicable accident protection and safety

regulations for electrical equipment.

protect the power supply
• Protect the power supply with a correctly rated

fuse if it is not protected by default.

• Use a shielded cable with an operating tem-

perature limit higher than +80 °C.

• Use a high quality electrical power supply (fil-

tered and regulated).

• Do not install the cable near high voltage or high

frequency cables.

• If a closed installation cannot be avoided,

maintain a minimum distance of 30 cm.

English

29

30

Assembling the female connector

1

2

3

4

Unscrew nut [1] of the cable gland.

Remove terminal block [3] from
housing [2].

Insert the cable into nut [1], through
seal [4], and into the cable gland and
finally through housing [2].

Connect the wires on terminal block
[3].

Position terminal block [3] in steps of
90° then put it back into housing [2],
pulling gently on the cable so that the
wires do not clutter the housing.

Tighten nut [1] of the cable gland.

6

5

Place seal [5] between the connector
and the fixed connector on the device
and then plug the 2508 connector into
the fixed connector.

Insert and then tighten screw [6] to
ensure tightness and correct electrical
contact.

Fig. 3: Assembling the female connector type 2508

(supplied)

English

31

1

2

3

4

1 : V+ (12-36 V DC)

2 : NPN transistor output

3 : 0 V DC

4 : PNP transistor output

Fig. 4: Pin assignment of the fixed connector, Hall

version

12-36 V DC

+ -

+

-

1

3

2

V+

0 V DC

Power supply

Terminal block of the 2508

Pulse input on external instrument

Fig. 5: NPN wiring of the Hall version

English