Installation of the fitting onto the pipe – Burkert Type 8020 User Manual
Page 7

24
Warning
risk of injury due to non-conforming commissioning.
Non-conforming commissioning may lead to injuries and
damage the device and its surroundings.
• Before commissioning, make sure that the staff in
charge have read and fully understood the contents of
the manual.
• In particular, observe the safety recommendations and
intended use.
• The device / the installation must only be commis-
sioned by suitably trained staff.
note
risk of damage to the device due to the environment
• Protect this device against electromagnetic interfer-
ence, ultraviolet rays and, when installed outdoors, the
effects of the climatic conditions.
To make sure the device operates correctly, plug in
and tighten the connector.
English
25
installation of the fitting onto the
pipe
→
Choose a fitting appropriate to the velocity and the
flow rate of the fluid inside the pipe, refer to the graphs
at right (see Tab. 3). The graph is used to determine
the DN of the pipe and the fitting appropriate to the
application, according to the fluid velocity and the flow
rate.
→
Install the fitting on the pipe as described in the oper-
ating instructions of the fitting used.
English
25
26
Tab. 3:
Diagram flow/ fluid velocity/ DN of S020 fitting
0.1
0.3 0.5 1
3 5 10
0.01
0.02
0.05
0.1
0.2
0.5
1
2
5
10
20
50
100
200
m
3
/h
0.2
0.5
1
2
5
10
20
50
100
200
500
1000
2000
3000
l/min
0.3 0.5 1
3 5 10
30
m/s
fps
US gpm
0.05
0.1
0.2
0.5
1
2
5
10
20
50
100
200
500
1000
DN 65
DN 50 (DN65)*
DN 40 (DN50)*
DN 32 (DN40)*
DN 25 (DN32)*
DN 20 (DN25)*
500
1000
2000
2000
5000
10000
5000
20000
5000
10000
20000
30000
50000
100000
DN 400
DN 350
DN 300
DN 250
DN 200
DN 150
DN 125
DN 100
DN 80
Flow rate
Fluid velocity
Example
1)
English
27
Example:
• Specification:
- nominal flow: 10 m
3
/h,
- optimal flow velocity: between 2 and 3 m/s
• Solution: intersection between flow rate and flow velocity
in the graph gives the appropriate pipe diameter, DN40
(or DN50 for the asterisked fittings).
* For the fittings:
• with external thread connections acc. to SMS 1145,
• with welding end connections acc. to SMS 3008,
BS 4825 / ASME BPE or DIN 11850 Rg 2,
• Clamp connections acc. to SMS 3017/ ISO 2852, BS
4825 / ASME BPE or DIN 32676
1)
The device cannot be installed on the DN20 fittings listed
above.
English
27