Burkert Type 6606 User Manual
Page 4
12
sysTem descripTiOn
6.
intended application area
6.1.
The solenoid valves Type 6126 and Type 6606 are designed
for use in analytical, medical and laboratory technology. They
can be used primarily for the metering, filling, mixing and
distribution of small amounts.
general description
6.2.
The solenoid valves Type 6126 and Type 6606 are direct-
acting rocker solenoid valves. They have a minimum dead
volume and a non-split, easy-to-rinse inner contour. The
medium comes into contact with the body material and the
seals only. The heat transfer into the medium is minimal, as
the body is also separated from the coil by a stainless steel
plate.
Modularity
The valve is modular in design and can be supplied with dif-
ferent port connections according to the application case. It
can be used individually and also in blocks.
english
13
7. Technical daTa
Operating conditions
7.1.
Warning!
Risk of injury
Malfunction if used outside!
Do not use Type 6126 and Type 6606 outdoors and avoid
•
heat sources which may cause the allowable temperature
range to be exceeded.
Allowable temperatures
Ambient temperature:
-10 - +55 °C
Medium temperature:
0 - +50 °C
Media:
neutral and aggressive, gaseous and
liquid media which do not attack body
and seal materials
see Bürkert Chemical Resistance Chart
(
www.burkert.com
)
Protection class: IP54 with flying leads or cable plug
IP40 with rectangular plug
english
13
14
conformity
7.2.
CE mark conforms to EMC Directive 2004/108/EEC
(only if cables, plugs and sockets connected correctly).
mechanical data
7.3.
Dimensions
See data sheet
Body material
PPS, PVDF, ETFE, PEEK
Brass, stainless steel
Seal material
FKM, FFKM, EPDM
english
15
7.4. pneumatic data
Circuit functions
A
A
P/NC
2/2-way valve, direct-acting,
normally closed
B
B
P/NO
2/2-way valve, direct-acting,
normally open
T
A in/out
P/NC R/NO
3/2-way valve, direct-acting,
universal function
Note the information specified on the label for
voltage, type of current, and pressure.
Pressure range
Vac ... 2 bar
Port connections
Threaded connection G1/8, M5
NPT1/8 or UNF1/4-28
Flange connection
Tube hose connection
english
15
Type 6126 / 6606