Model sla7840 – Brooks Instrument SLA7840 User Manual
Page 13

2-1
Model SLA7840
Section 2 Installation
Installation and Operation Manual
X-PR-SLA7800-RT-eng
Part Number: 541B048AAG
August, 2009
2-1 General
This section provides installation instructions for the Brooks Model SLA7840
Remote Transducer. Figures 1-1 and 1-2 show the Model SLA7840
dimensions, gas connections and electrical connection locations for an RT.
2-2 Receipt of Equipment
When the instrument is received, the outside packing case should be
checked for damage incurred during shipment. If the packing case is
damaged, the local carrier should be notified at once regarding his liability.
A report should be submitted to your nearest Product Service Department.
Brooks Instrument
407 W. Vine Street
P.O. Box 903
Hatfield, PA 19440 USA
Toll Free (888) 554 FLOW (3569)
Tel (215) 362 3700
Fax (215) 362 3745
E-mail: [email protected]
www.BrooksInstrument.com
Brooks Instrument
Brooks Instrument
Neonstraat 3
1-4-4 Kitasuna Koto-Ku
6718 WX Ede, Netherlands
Tokyo, 136-0073 Japan
P.O. Box 428
Tel +81 (0) 3 5633 7100
6710 BK Ede, Netherlands
Fax +81 (0) 3 5633 7101
Tel +31 (0) 318 549 300
Email: [email protected]
Fax +31 (0) 318 549 309
E-mail: [email protected]
Remove the envelope containing the packing list. Outside of your clean area,
carefully remove the equipment from the packing case. Make sure spare
parts are not discarded with the packing material. Inspect for damaged or
missing parts.
This device has been assembled, calibrated and double-vacuum bagged in
a Class 100 clean room. In your semi-clean area,remove the outer bag
only. Pass your Model SLA7840 Ultra-Clean Metal Seal Remote
Transducer Controller into your clean area. Remove the second clean
room compatible bag only when the equipment is ready to be tested and/or
installed in your clean system.
2-3 Recommended Storage Practice
If intermediate or long-term storage of equipment is required, it is
recommended that the equipment be stored in accordance with the
following:
a. In the original vacuum bag and shipping container.
b. In a sheltered area with the following conditions:
1. Ambient temperature 21°C (70°F) nominal, 40°C (104°F) maximum
and 7°C (45°F) minimum.
2. Relative humidity 45% nominal, 60% maximum and 25% minimum.