14 - troubleshooting – CEMB USA DWA 1000XLB User Manual
Page 54

I 0645
55
GB
14 - TROUBLESHOOTING
14.1 - Malfunctions
[L] = Linux models
[W] = Windows models
[R] = Radio models
[C] = Cable models
The computer does not turn on
No power supply
Check the mains socket and the connections
Press the computer power-on button
see par. 1.1
[L] Reset button not pressed
Press the Reset button
see paragraph 1.1
Display off
Turn on the display (button on display)
The measuring head does not turn on or turns off unexpectedly
[R] Batteries fl at
Charge the batteries
[C] No power supply
Check the mains socket and the connections
[R] Battery recharging does not start
No power supply
Check the mains socket and the connections
Press the battery recharge button
see paragraph. 1.3
[R] The battery life is too short
Low ambient temperature
Normal phenomenon
Recharging not complete
Recharge the batteries for at least 5 hours
The battery has come to the end
of its lifetime
Replace the battery