Calibrate the current value (0x2b), Save the calibration data to eeprom (0x2d, Setting calibration information (0x2e) – B&K Precision 1785B - Manual User Manual
Page 28: Reading calibration information (0x2f)
![background image](/manuals/320562/28/background.png)
28
12. Calibrate the current value (0x2B)
1
st
byte
Start bit(0xAA)
2
nd
byte
Address(0x00-0xFE)
3
rd
byte
Command(0x2B)
4
th
byte
Calibrated current points( point 1-2)
5
th
to 25
th
byte
System reserve
26
th
byte
Check sum
13. Sending the actual output current to calibration program (0x2C)
1
st
byte
Start bit (0xAA)
2
nd
byte
Address(0x00~0xFE)
3
rd
byte
Command(0x2C)
4
th
byte
The lower byte of the present current value
5
th
byte
The higher byte of the present current value
6
th
to 25
th
byte
System reserve
26
th
byte
Check sum
14. Save the calibration data to EEPROM(0x2D)
1
st
byte
Start bit(0xAA)
2
nd
byte
Address (0x00~0xFE)
3
rd
byte
Command(0x2D)
4
th
to 25
th
byte
System reserve
26
th
byte
Check sum
15. Setting calibration information (0x2E)
1
st
byte
Start bit (0xAA)
2
nd
byte
Address (0x00~0xFE)
3
rd
byte
Command(0x2E)
4
th
to 23
rd
byte
Calibration information(ASIC code)
24
th
byte
System reserve
25
th
byte
System reserve
26
th
byte
Check sum
16. Reading calibration information (0x2F)
1
st
byte
Start bit (0xAA)
2
nd
byte
Address (0x00~0xFE)
3
rd
byte
Command(0x2F)
4
th
to 23
rd
byte
Calibration information(ASCII code)
24
th
byte
System reserve
25
th
byte
System reserve
26
th
byte
Check sum
Note: To calibrate the 2 points of the current value sequentially.