Calibrate the current value (0x2b), Save the calibration data to eeprom (0x2d, Setting calibration information (0x2e) – B&K Precision 1785B - Manual User Manual

Page 28: Reading calibration information (0x2f)

background image

28

12. Calibrate the current value (0x2B)

1

st

byte

Start bit(0xAA)

2

nd

byte

Address(0x00-0xFE)

3

rd

byte

Command(0x2B)

4

th

byte

Calibrated current points( point 1-2)

5

th

to 25

th

byte

System reserve

26

th

byte

Check sum







13. Sending the actual output current to calibration program (0x2C)

1

st

byte

Start bit (0xAA)

2

nd

byte

Address(0x00~0xFE)

3

rd

byte

Command(0x2C)

4

th

byte

The lower byte of the present current value

5

th

byte

The higher byte of the present current value

6

th

to 25

th

byte

System reserve

26

th

byte

Check sum


14. Save the calibration data to EEPROM(0x2D)

1

st

byte

Start bit(0xAA)

2

nd

byte

Address (0x00~0xFE)

3

rd

byte

Command(0x2D)

4

th

to 25

th

byte

System reserve

26

th

byte

Check sum



15. Setting calibration information (0x2E)

1

st

byte

Start bit (0xAA)

2

nd

byte

Address (0x00~0xFE)

3

rd

byte

Command(0x2E)

4

th

to 23

rd

byte

Calibration information(ASIC code)

24

th

byte

System reserve

25

th

byte

System reserve

26

th

byte

Check sum


16. Reading calibration information (0x2F)

1

st

byte

Start bit (0xAA)

2

nd

byte

Address (0x00~0xFE)

3

rd

byte

Command(0x2F)

4

th

to 23

rd

byte

Calibration information(ASCII code)

24

th

byte

System reserve

25

th

byte

System reserve

26

th

byte

Check sum

Note: To calibrate the 2 points of the current value sequentially.