4.4 Communication protocol

1Setting the remote control mode (0x20)

1

st

byte

Start bit( 0xAA )

2

nd

byte

Address(0x00~0xFE)

3

rd

byte

Command (0x20)

4

th

byte

Operation mode(0 represent front panel operation mode, 1
represent remote operation mode)

5

th

to 25

th

byte

System reserve

26

th

byte

Check sum







2Setting the output state ON/OFF (0x21)

1

st

byte

Start bit (0xAA )

2

nd

byte

Address(0x00~0xFE)

3

rd

byte

Command (0x21)

4

th

byte

Output state(0 is OFF,1 is ON)

5

th

to 25

th

byte

System reserve

26

th

byte

Check sum



3Setting the maximum output voltage (0x22)

1

st

byte

Start bit (0xAA )

2

nd

byte

Address(0x00~0xFE)

3

rd

byte

Command (0x22)

4

th

byte

The lowest byte of voltage upper limit (1 represents 1 mV)

5

th

byte

The lower byte of voltage upper limit

6

th

byte

The higher byte of voltage upper limit

7

th

byte

The highest byte of voltage upper limit

8

th

to 25

th

byte

System reserve

26

th

byte

Check sum










Note:

Suppose you want to set the maximum voltage to 16.23 V. Since 1

represents 1mV, therefore 16.23 V translates to 16,230 in decimal. With 4 bytes in
Hex, that would be 0x0003F66. Since the bytes are ordered in little-endian format,
0x66 would be the 3

rd

byte, 0x3F the 4

th

byte, 0x00 as 5

th

byte, and 0x00 as the 6

th

byte.