beautypg.com

8 . entering the calibration mode (0x27) – B&K Precision 1785B - Manual User Manual

Page 26

background image

26

9

th

byte

Byte 3 of measured output voltage

10

th

byte

Power supply’s state

11

th

byte

Byte 0 of current value setting (current limit value)

12

th

byte

Byte 1 of current value setting

13

th

byte

Byte 0 of the maximum voltage setting

14

th

byte

Byte 1 of the maximum voltage setting

15

th

byte

Byte 2 of the maximum voltage setting

16

th

byte

Byte 3 of the maximum voltage setting

17

th

byte

Byte 0 of voltage value setting

18

th

byte

Byte 1 of voltage value setting

19

th

byte

Byte 2 of voltage value setting

20

th

byte

Byte 3 of voltage value setting

21

st

to 25

th

byte

System reserve

26

th

byte

Check sum



















8
. Entering the calibration mode(0x27)

1

st

byte

Start bit(0xAA)

2

nd

byte

Address(0x00~0xFE)

3

rd

byte

Command(0x27)

4

th

byte

Calibration protection state

5

th

byte

Calibration password(0x28)

6

th

byte

Calibration password(0x01)

7

th

to 25

th

byte

System reserve

26

th

byte

Check sum



Note:

1. We use 4 bytes to represent the maximum voltage value as follows:

Byte 3

Byte 2

Byte1

Byte0

2. We use 1 byte to represent power supply’s state. Each bit is defined as

follows:

From higher bit to lower bit

7

6

5

4

3

2

1

0

0 bitThe output state, 0 is OFF, 1 is ON.
1 bit
Over heat protection, 0 is normal, 1 is abnormal.
2
3 bit: The output mode, 1 is CV mode, 2 is CC mode,3 is Unreg mode.
4
56 bitThe fan speed, 0 is stop, 5 is the maximum fan speed.

7 bitOperation state, 0 is front panel operation mode, 1 is remote control

mode.

3. The frame format is the same as above

This manual is related to the following products: