Specifications

Specifications

Specifications

18

Integration timer

opped automatically using the timer setting. 0000h00m00s to

Integration can be stopp
10000h00m00s

10000h00 00

Count over

If the integration time reaches the maximum integration time (10000 hours),
or if the integration value reaches max/min display integration value

*1

,

the elapsed time and integration value is saved and the operation is stopped.

*1: WP

: ±999999 MWh

q

: ±999999 MAh

WS

: ±999999 MVAh

WQ

: ±999999 Mvarh

Accuracy

±(Normal measurement accuracy + 0.02% of reading)

Timer accuracy

±0.02% of reading

Harmonic Measurement

(Option)

Item

Specification

Measured source

All installed elements

Method

PLL synchronization method (without external sampling clock function)

Frequency range

Fundamental frequency of the PLL source is in the range of 0.5 Hz to 2.6 kHz.

PLL source

• Select the voltage or current of each input element or the external clock.
• If the /G6 option is selected, two PLL sources can be selected, and dual
harmonic measurement can be performed. If the /G5 option is selected,
one PLL source is selectable.
• Input level
15 V or more of range for voltage input.
50 mA or more of range for direct current input.
200 mV or more of range for external current sensor input.
50% or more of the measurement range rating for crest factor 3.
100% or more of the measurement range rating for crest factor 6.
20 Hz to 1 kHz for the 1 A or 2 A range of the 50 A input element.
• The frequency fi lter ON condition is the same as with frequency measurement.

FFT data length

1024 when the data update rate is 50 ms, 100 ms, or 200 ms
8192 when the data update rate is 500 m, 1 s, 2 s, 5 s, 10 s, or 20 s

Window function

Rectangular

Anti-aliasing fi lter

Set using a line fi lter

Sample rate, window width, and upper limit of the measured order

1024 FFT points (data update rate 50 ms, 100 ms, 200 ms)

Upper limit of measured order

Fundamental frequency Sampling rate Window width U, I, P, Ø, ØU, ØI or other measured values
15 Hz to 600 Hz

f*1024

1

500th order

100th order

600 Hz to 1200 Hz

f*512

2

255th order

100th order

1200 Hz to 2600 Hz

f*256

4

100th order

100th order

However, the maximum measured order is 100 at a date update rate of 50 ms.

8192 FFT points (data update rate 500 m, 1 s, 2 s, 5 s, 10 s, 20 s)

Upper limit of measured order

Fundamental frequency Sampling rate Window width U, I, P, Ø, ØU, ØI or other measured values
0.5 Hz to 1.5 Hz

f*8192

1

500th order

100th order

1.5 Hz to 5Hz

f*4096

2

500th order

100th order

5 Hz to 10 Hz

f*2048

4

500th order

100th order

10 Hz to 600 Hz

f*1024

8

500th order

100th order

600 Hz to 1200 Hz

f*512

16

255th order

100th order

1200 Hz to 2600 Hz

f*256

32

100th order

100th order

Item

Specification

Accuracy

Add the following accuracy to the normal measurement accuracy.

When the line fi lter is OFF
Frequency

Voltage

Current

Power

0.5 Hz

) f < 10 Hz

0.05% of reading
+ 0.25% of range

0.05% of reading
+ 0.25% of range

0.1% of reading
+ 0.5% of range

10 Hz

) f < 45 Hz

0.05% of reading
+ 0.25% of range

0.05% of reading
+ 0.25% of range

0.1% of reading
+ 0.5% of range

45 Hz

) f ) 66 Hz

0.05% of reading
+ 0.25% of range

0.05% of reading
+ 0.25% of range

0.1% of reading
+ 0.5% of range

66 Hz < f

) 440 Hz

0.05% of reading
+ 0.25% of range

0.05% of reading
+ 0.25% of range

0.1% of reading
+ 0.5% of range

440 Hz < f

) 1 kHz

0.05% of reading
+ 0.25% of range

0.05% of reading
+ 0.25% of range

0.1% of reading
+ 0.5% of range

1 kHz < f

) 10 kHz

0.5% of reading
+ 0.25% of range

0.5% of reading
+ 0.25% of range

1% of reading
+ 0.5% of range

10 kHz < f

) 100 kHz

0.5% of range

0.5% of range

1% of range

100 kHz < f

) 260 kHz

1% of range

1% of range

2% of range

• When the line fi lter is ON

Add the accuracy of the line fi lter to the accuracy of when the line fi lter is OFF

All the items below apply to any of the tables.
• When the crest factor is set to 3
• When

λ (power factor) = 1

• Power fi gures that exceed 2.6 kHz are reference values.
• For the voltage range, add the following values.

Voltage accuracy: 25 mV
Power accuracy: (25 mV/voltage range rating) × 100% of range

• For the direct current input range, add the following values.

5 A element
Current accuracy: 50 μA
Power accuracy: (50 μA/current range rating) × 100% of range
50 A element
Current accuracy: 4 mA
Power accuracy: (4 mA/current range rating) × 100% of range

• For the external current sensor range, add the following values.

Current accuracy: 2 mV
Power accuracy: (2 mV/external current sensor range rating) × 100% of range

• Add (n/500)% of reading to the n-th component of the voltage and current, and add (n/250)% of reading to

the n-th component of the power.

• Accuracy when the crest factor is 6: Same as when the range is doubled for crest factor 3
• The guaranteed accuracy range by frequency and voltage/current is the same as the guaranteed range of

normal measurement.

• The adjacent orders of the input order may be affected by the side rope.
• For n-th order component input when the PLL source frequency is 2 Hz or more, add ({n/(m+1)}/50)% of (the

n-th order reading) to the (n+m)th order and (n-m)th order of the voltage and current, and add ({n/(m+1)}/25)%
of (the n-th order reading) to the (n+m)th order and (n-m)th order of the power.

• For n-th order component input when the PLL source frequency is less than 2 Hz, add ({n/(m+1)}/20)% of (the

n-th order reading) to the (n+m)th order and (n-m)th order of the voltage and current, and add ({n/(m+1)}/10)%
of (the n-th order reading) to the (n+m)th order and (n-m)th order of the power.

Motor Evaluation Function

(Option)

Item

Specification

Input terminal

Torque, speed (A, B, Z)

Input resistance

Approximately 1 M

1

Input connector type

Insulated BNC

Analog Input (Speed is input to the A terminal)

Item

Specification

Range

1 V, 2 V, 5 V, 10 V, 20 V

Input range

±110%

Line fi lter

OFF, 100, 1 kHz

Continuous maximum allowable input ±22 V
Maximum common mode voltage

±42 Vpeak

Sampling rate

Approximately 200 kS/s

Resolution

16-bit

Accuracy

±(0.05% of reading + 0.05% of range)

Temperature coeffi cient

±0.03% of range/°C

Pulse Input

Speed is input to the A terminal if the direction is not detected. If the direction is detected, the A and B phases
of the rotary encoder are input to the A and B terminals. The Z phase is input to the Z terminal of the rotary
encoder for electric angle measurement.

g

Item

Specification

Input range

±12 Vpeak

Frequency measurement range

2 Hz to 1 MHz

Maximum common mode voltage

±42 Vpeak

Accuracy

±(0.05 + f/500)% of reading ±1 mHz

Rise of the Z terminal input and electric angle measurement start time

Within 500 ns

Detection level

H level: Approximately 2 V or more
L level: Approximately 0.8 V or less

Pulse width

500 ns or more

Harmonic measurement option (/G5 or /G6) is required for electric angle measurement.

Auxiliary Input

(Option)

Item

Specification

Input terminal

AUX1/AUX2

Input type

Analog

Input resistance

Approximately 1 M

1

Input connector type

Insulated BNC

Range

50 m, 100 m, 200 m, 500 m, 1 V, 2 V, 5 V, 10 V, 20 V

Input range

±110%

Line fi lter

OFF/100 Hz/1 kHz

Continuous maximum allowable input

±22 V

Common mode voltage

±42 V

Sampling rate

Approximately 200 kS/s

Resolution

16-bit

Accuracy

±(0.05% of reading + 0.05% of range)
• Add 20 μV/°C to the change in temperature after zero level compensation

or range change.

Temperature coeffi cient

±0.03% of range/°C

DA Output and Remote Control

(Option)

DA Output

Item

Specification

D/A conversion resolution 16-bit
Output voltage

±5 V FS (max. approximately ±7.5 V) relative to each rated value

Update rate

Same as the data update rate

Output

20 channels (Output parameter can be set for each channel)

Accuracy

± (Accuracy of each measurement function +0.1% of FS) FS=5 V

Minimum load

100 k

1

Temperature coeffi cient

±0.05% of FS/°C

Continuous maximum common mode voltage

±42 Vpeak or less

Remote Control

Item

Specification

Signal

EXT START, EXT STOP,

P

P EXT RESET, INTEG BUSY, EXT HOLD, EXT SINGLE, EXT PRINT

Input level

0 to 5 V

Calculation and Event Function

Item

Specification

User-defi ned function

Compute the numerical data (up to 20 equations) with a combination of
measurement function symbols and operators.

Effi ciency calculation

Up to 4 effi ciencies can be displayed by setting measurement parameters
for the effi ciency equations.

User-defi ned event

Event: Set conditions for measured values.
The functions triggered by the event are Auto Print, Store, and DA Output.