Flow transmitter, Technical data – Atec Flexim-G601 User Manual

Page 6

background image

6

TSFLUXUS_G601V1-5-1US_Lus, 2014-02-25

FLUXUS® G601

Technical Specification

Flow Transmitter

Technical Data

FLUXUS

G601

design

portable

measurement
measurement principle

transit time difference correlation principle

flow velocity

0.03 to 115 ft/s, depending on pipe diameter

repeatability

0.15 % of reading ±0.03 ft/s

medium

all acoustically conductive gases,
e.g. nitrogen, air, oxygen, hydrogen, argon, helium, ethylene, propane

temperature compensation

corresponding to the recommendations in ANSI/ASME MFC-5.1-2011

accuracy
volumetric flow rate

± 1 to 3 % of reading ±0.03 ft/s depending on application
± 0.5 % of reading ±0.03 ft/s with field calibration

flow transmitter
power supply

100 to 240 V/50 to 60 Hz (power supply unit),
10.5 to 15 V DC (socket at transmitter),
integrated battery

battery

Li-Ion, 7.2 V/4.5 Ah
operating time (without outputs, inputs and backlight): > 14 h

power consumption

< 6 W

number of flow measuring
channels

2

signal attenuation

0 to 100 s, adjustable

measuring cycle (1 channel) 100 to 1000 Hz
response time

1 s (1 channel), option: 70 ms

housing material

PA, TPE, AutoTex, stainless steel

degree of protection

NEMA 4

dimensions

see dimensional drawing

weight

4.2 lb

fixation

QuickFix pipe mounting fixture

ambient temperature

14 to 140 °F

display

2 x 16 characters, dot matrix, backlight

menu language

English, German, French, Dutch, Spanish

measuring functions
physical quantities

operating volumetric flow rate, standard volumetric flow rate, mass flow rate, flow velocity

totalizer

volume, mass

calculation functions

average, difference, sum

diagnostic functions

sound speed, signal amplitude, SNR, SCNR, standard deviation of amplitudes and transit times

data logger
loggable values

all physical quantities, totalized values and diagnostic values

capacity

> 100 000 measured values