Clamp-on temperature probe (optional), Technical data connection, Cable – Atec Flexim-G601 User Manual
Page 26
26
TSFLUXUS_G601V1-5-1US_Lus, 2014-02-25
FLUXUS® G601
Technical Specification
Clamp-on Temperature Probe (optional)
Technical Data
Connection
Temperature Probe
Connector
Cable
technical type
PT13N
PT13N
PT13F
PT13F
order code
670413-1
670412-1
670413-2
670412-2
design
short response time
type
Pt1000
2x Pt1000 matched
according to EN 1434-1
Pt1000
2x Pt1000 matched
according to EN 1434-1
connection
4-wire
4-wire
measuring range
°F
-22 to +482
-58 to +482
accuracy T
±(0.27 °F + 2
.
10
-3
.
(|T [°F]| - 32 °F)), class A
±(0.27 °F + 2
.
10
-3
.
(|T [°F]| - 32 °F)), class A
accuracy ∆T
-
≤ 0.1 K, (3 K < ∆T < 6 K),
more corresponding to
EN 1434-1
-
≤ 0.1 K, (3 K < ∆T < 6 K),
more corresponding to
EN 1434-1
response time
s
50
8
housing
aluminum
PEEK, stainless steel 304, copper
degree of protection
NEMA 4
NEMA 4
weight (without con-
nector)
lb
0.6
1.1
0.7
1.4
fixation
clamp-on
clamp-on
accessories
thermal conductivity
paste 392 °F
x
x
thermal conductivity
foil 482 °F
x
x
plastic protection plate,
insulation foam
-
x
dimensions
length l
in
0.59
0.55
width b
in
0.59
1.18
height h
in
0.79
1.06
dimensional drawing
A
B
pin
cable of temperature probe
extension cable
1
white/blue
blue
2
red/blue
gray
3, 4, 5
not connected
6
red
red
7
white
white
8
not connected
cable of temperature probe
extension cable
type
4 x 0.25 mm² black or white
LIYCY 8 x 0.14 mm² gray
standard length
ft
9
16/32/82
max. length
ft
-
656
cable jacket
PTFE
PVC
h
l
extension cable
h
l
A
B
red/blue
red
white/blue
white
8
1
2
3
4
5
6
7