beautypg.com

Technical data – Atec Flexim-G601 User Manual

Page 16

background image

16

TSFLUXUS_G601V1-5-1US_Lus, 2014-02-25

FLUXUS® G601

Technical Specification

Technical Data

Shear Wave Transducers

technical type

GDG1NZ7

GDK1NZ7

order code

GSG-NNNNL

GSK-NNNNL

transducer frequency

MHz 0.2

0.5

medium pressure

1

min. extended

psi

metal pipe: 290

metal pipe: 290

min.

psi

metal pipe: 435
plastic pipe: 15

metal pipe: 435
plastic pipe: 15

inner pipe diameter d

2

min. extended

in

9.8

2.8

min. recommended

in

15

3.1

max. recommended

in

31.9

19.7

max. extended

in

43.3

28.3

pipe wall thickness
min.

in

0.55

0.2

max.

in

-

-

material
housing

PEEK with stainless steel
cap 304

PEEK with stainless steel
cap 304

contact surface

PEEK

PEEK

degree of protection

NEMA 6

NEMA 6

transducer cable
type

1699

1699

length

ft

16

16

dimensions
length l

in

5.1

4.98

width b

in

2.01

2.01

height h

in

2.64

2.66

dimensional drawing

ambient temperature
min.

°F

-40

-40

max.

°F

+266

+266

temperature
compensation

x

x

1

depending on application, typical absolute value for natural gas, nitrogen, compressed air

2

shear wave transducer:
typical values for natural gas, nitrogen, oxygen, pipe diameters for other gases on request
pipe diameter min. recommended/max. recommended/max. extended: in diagonal arrangement and for a flow velocity of 49 ft/s

l

h

b

l

h

b