beautypg.com

Atec Flexim-G601 User Manual

Page 19

background image

TSFLUXUS_G601V1-5-1US_Lus, 2014-02-25

19

Technical Specification

FLUXUS® G601

Lamb Wave Transducers

technical type

GRG1NC3

GRH1NC3

GRK1NC3

order code

GLG-NNNNL

GLH-NNNNL

GLK-NNNNL

transducer frequency

MHz 0.2

0.3

0.5

medium pressure

1

min. extended

psi

metal pipe: 145

metal pipe: 145

metal pipe:
145 (d > 4.7 in)
73 (d < 4.7 in)

min.

psi

metal pipe: 218
plastic pipe: 15

metal pipe: 218
plastic pipe: 15

metal pipe:
218 (d > 4.7 in)
145 (d < 4.7 in)
plastic pipe: 15

inner pipe diameter d

2

min. extended

in

7.5

4.7

2.4

min. recommended

in

8.7

5.5

3.1

max. recommended

in

35.4

23.6

11.8

max. extended

in

63

39.4

19.7

pipe wall thickness
min.

in

0.43

0.28

0.16

max.

in

0.91

0.59

0.35

material
housing

PPSU with stainless steel
cap 304

PPSU with stainless steel
cap 304

PPSU with stainless steel
cap 304

contact surface

PPSU

PPSU

PPSU

degree of protection

NEMA 4

NEMA 4

NEMA 4

transducer cable
type

1699

1699

1699

length

ft

16

16

16

dimensions
length l

in

5.06

5.06

5.06

width b

in

2.01

2.01

2.01

height h

in

2.66

2.66

2.66

dimensional drawing

ambient temperature
min.

°F

-40

-40

-40

max.

°F

+338

+338

+338

temperature
compensation

x

x

x

1

depending on application, typical absolute value for natural gas, nitrogen, compressed air

2

Lamb wave transducer:
typical values for natural gas, nitrogen, oxygen, pipe diameters for other gases on request
pipe diameter min. recommended/max. recommended: in reflect arrangement and for a flow velocity of 49 ft/s
pipe diameter max. extended: in diagonal arrangement and for a flow velocity of 82 ft/s

l

h

b

l

h

b

l

h

b