Transducers, Transducer selection – Atec Flexim-G601 User Manual

Page 12

background image

12

TSFLUXUS_G601V1-5-1US_Lus, 2014-02-25

FLUXUS® G601

Technical Specification

Transducers

Transducer Selection

Step 1a
Select a Lamb wave transducer:

Step 1b
If the pipe wall thickness is not in the range of the Lamb wave transducers, select a shear wave transducer:

Step 2
inner pipe diameter d dependent on the flow velocity v of the medium in the pipe
The transducers are selected from the characteristics (see next page). Lamb wave transducers are selected

from the left column, shear wave transducers from the right column.
Lamb wave transducers: If the values d and v are not in the range, the diagonal arrangement with 1 sound

path may be used, i.e. the same characteristics can be used with doubling the inner pipe diameter. If the val-

ues are still not in the range, shear waves transducers regarding the pipe wall thickness have to be selected

in step 1b.

transducer order code

GLG

0.43

0.91

GLH

0.28

0.59

GLK

0.16

0.35

GLM

0.08

0.2

GLP

0.04 to 0.12

GLQ

0.02 to 0.04

0.2

0.4

0.6

0.8

1

pipe wall thickness [in]

transducer order code

GSG

0.55

0.91

GSK

0.2

0.35

GSM

0.1

0.2

GSP

0.06 0.12

GSQ

0.04

0.06

0.2

0.4

0.6

0.8

1

pipe wall thickness [in]

recommended

possible