TSFLUXUS_G601V1-5-1US_Lus, 2014-02-25
FLUXUS® G601
Technical Specification
Step 3
min. medium pressure
Example
Step 4
for the characters 4
to
11 of the transducer order code (ambient temperature, explosion protection, connection
system, extension cable) see page 15
Step 5
for the technical data of the selected transducer see page 16 et seqq.
Lamb wave transducer
shear wave transducer
transducer
order code
medium pressure
1
[psi]
transducer
order code
medium pressure
1
[psi]
metal pipe
plastic pipe
metal pipe
plastic pipe
min.
min. extended
min.
min.
min. extended
min.
GLG
218
145
15
GSG
435
290
15
GLH
218
145
15
GSK
435
290
15
GLK
218 (d > 4.7 in)
145 (d < 4.7 in)
145 (d > 4.7 in)
73 (d < 4.7 in)
15
GSM
435
290
15
GLM
145 (d > 2.4 in)
73 (d < 2.4 in)
-
15
GSP
435
290
15
GLP
145 (d > 1.4 in)
73 (d < 1.4 in)
-
15
GSQ
435
290
15
GLQ
145 (d > 0.59 in)
73 (d < 0.59 in)
-
15
1
depending on application, typical absolute value for natural gas, nitrogen, compressed air
d = inner pipe diameter
step
1
pipe wall thickness
in
0.47
0.47
0.47
1.2
selected transducer
GLG or GLH
GLG or GLH
GLG or GLH
GS
2
inner pipe diameter
in
31.5
23.6
31.5
11.8
max. flow velocity
ft/s
49
49
98
49
selected transducer
GLG
GLG or GLH
values not in the range
of the characteristics,
but by using direct
mode, the inner pipe
diameter in the cha-
racteristics is doubled:
GLG
GSK
3
min. medium pressure
psi
247
247
247
508
selected transducer
GLG
GLG or GLH
influence of acoustic
noise is reduced with
increased transducer
frequency, thus
recommended:
GLH
GLG
GSK