Atec Rohde-Schwarz-ESIB Series User Manual
Page 9
EMI Test Receivers ¸ESIB
9
¸ESIB 7
¸ESIB 26
¸ESIB 40
Preselector (receiver mode)
Filters
Frequency range
Bandwidth (–6 dB)
1
<150 kHz
230 kHz
fi xed
2
150 kHz to 2 MHz
2.6 MHz
fi xed
3
2 MHz to 8 MHz
1.9 MHz
tracking
4
8 MHz to 25 MHz
5.6 MHz
tracking
5
25 MHz to 80 MHz
15 MHz
tracking
6
80 MHz to 200 MHz
40 MHz
tracking
7
200 MHz to 500 MHz
85 MHz
tracking
8
500 MHz to 1000 MHz
104 MHz
tracking
9
1 GHz to 7 GHz
highpass fi lter
fi xed
10
–
7 GHz to 26.5 GHz
YIG fi lter 7 GHz to 40 GHz
YIG fi lter
–
bandwidth (–3 dB): 35 MHz + f/1000
Preamplifi er (1 kHz to 7 GHz)
selectable, between preselector and 1st mixer, gain 20 dB
IF bandwidths (receiver and analyzer mode)
6 dB bandwidths
10 Hz, 100 Hz, 200 Hz, 1 kHz, 9 kHz, 10 kHz, 100 kHz, 120 kHz, 1 MHz
3)
, 10 MHz
Bandwidth error
RBW
≤
1 MHz
Shape factor B
60 dB
: B
6 dB
RBW
≤
1 kHz
RBW >1 kHz
<10 %
<5
<10
Resolution bandwidths (analyzer mode)
3 dB bandwidths
1 Hz to 10 MHz, in steps of 1/2/3/5
Bandwidth error
RBW
≤
3 MHz
RBW = 5 MHz
RBW = 10 MHz
Shape factor B
60 dB
: B
6 dB
RBW <1 kHz
RBW = 1 kHz to 2 MHz
RBW >2 MHz
<10 %
<15 %
+25 %, –10 %
<6
<12
<7
Video bandwidths
1 Hz to 10 MHz, in steps of 1/2/3/5
FFT fi lter
3 dB bandwidths
Bandwidth error, nominal
Shape factor B
60 dB
: B
3 dB
, nominal
Display range for frequency axis
Additional level error (reference: RBW = 5 kHz)
Max. display range
Inherent spurious response
1 Hz to 1 kHz, in steps of 1/2/3/5
2 %
2.5
min. 25 × RBW, max. 100000 × RBW or 2 MHz
<1 dB
100 dB
<–100 dBm
Level
Display range
displayed noise fl oor to 137 dBμV
Max. input level
Input 1
20 Hz to 7 GHz
20 Hz to 26.5 GHz
20 Hz to 40 GHz
RF attenuation 0 dB
DC voltage
0 V
Sinewave AC voltage
127 dBμV (= 0.3 W)
Pulse spectral density
97 dB(μV/MHz)
RF attenuation
≥
10 dB
DC voltage
0 V
Sinewave AC voltage
137 dBμV (= 1 W)
Max. pulse voltage (10 μs)
150 V
50 V
50 V
Max. pulse energy (10 μs)
1 mWs
0.5 mWs
0.5 mWs
ESIB_dat_en.indd 9