EMI Test Receivers ¸ESIB
¸ESIB 7
¸ESIB 26
¸ESIB 40
Probe power supply
+15 V DC, –12.6 V DC and ground, max. 150 mA
Power supply and coding connector for antennas etc
(antenna code)
12-contact Tuchel
Supply voltages
±10 V, max. 100 mA, ground
AF output
Z
out
= 10
Ω
, jack plug
Open-circuit voltage
up to 1.5 V, adjustable
Inputs and outputs (rear panel)
IF 21.4 MHz
Level
Z
out
= 50
Ω
, BNC female, bandwidth >1 kHz or IF or resolution bandwidth
0 dBm at reference level, mixer level >–60 dBm
Video output
Voltage (resolution bandwidth
≥
1 kHz)
Z
out
= 50
Ω
, BNC female
0 V to 1 V, full scale (open-circuit voltage), logarithmic scaling
Reference frequency
Output, usable as input
Output frequency
Level
Input
Required level
BNC female
10 MHz
10 dBm nominal
1 MHz to 16 MHz, in steps of 1 MHz
>0 dBm into 50
Ω
Sweep output
BNC female, 0 V to +10 V in sweep range
Power supply connector for noise source
BNC female, 0 V and 28 V, switch-selected
External trigger/gate input
Voltage
BNC female, >10 k
Ω
–5 V to +5 V, adjustable
IEC/IEEE bus remote control
Command set
Connector
Interface functions
interface in line with IEC 625-2 (IEEE 488.2)
SCPI 1994.0
24-contact Amphenol female
SH1, AH1, T6, L4, SR1, RL1, PP1, DC1, DT1, C11
Serial interface
RS-232-C (COM1 and COM2), 9-contact female connectors
Mouse interface
PS/2-compatible
Printer interface
parallel (Centronics-compatible) or serial (RS-232-C)
Keyboard connector
5-contact DIN female for MF2 keyboard
User interface
25-contact Cannon female
Connector for external monitor (VGA)
15-contact female
General data
Display
Resolution
Pixel error rate
24 cm LC color display (9.5“)
640 × 480 pixels (VGA resolution)
<2 × 10
–5
Mass memory
1.44 Mbyte 3½“ disk drive, hard disk
Temperature ranges
Operating temperature range
Permissible temperature range
Storage temperature range
+5 °C to +40 °C
0 °C to +50 °C
–40 °C to +70 °C
Environmental conditions
+40 °C at 95 % relative humidity (IEC 68-2-3)
Mechanical stress
Sinewave vibration
Random vibration
Shock
5 Hz to 150 Hz, max. 2 g at 55 Hz, 0.5 g from 55 Hz to 150 Hz; in line with IEC 68-2-6,
IEC 68-2-3, IEC 1010-1, MIL-T-28800D, class 5
10 Hz to 300 Hz, acceleration 1.2 g RMS
40 g shock spectrum, in line with MIL-STD-810C and MIL-T-28800D, classes 3 and 5
Recommended calibration interval
1 year (2 years for operation with external reference)
EMC
in line with CISPR 11/EN 55011 group 1 class B; in line with IEC/EN 61326, emission: class B
(residential environment), immunity: industrial environment (including operating frequency);
in line with CISPR 16-1-1
Power supply
AC supply
200 V to 240 V: 50 Hz to 60 Hz, 100 V to 120 V: 50 Hz to 400 Hz,
safety class in line with VDE 411
Power consumption
195 VA
230 VA
Safety
in line with EN 61010-1, UL 3111-1, CSA C22.2 No. 1010-1, IEC 1010-1
Test mark
VDE, GS, UL, cUL
ESIB_dat_en.indd 14
04.01.2006 13:27:24