10

EMI Test Receivers ¸ESIB

¸ESIB 7

¸ESIB 26

¸ESIB 40

Input 2 (receiver mode)

20 Hz to 1 GHz

DC voltage

DC coupling

0 V

AC coupling

50 V

RF attenuation 0 dB

Sinewave AC voltage

127 dBμV (= 0.3 W)

Pulse spectral density

97 dB(μV/MHz)

RF attenuation

10 dB

Sinewave AC voltage

137 dBμV (= 1 W)

Max. pulse voltage (10 μs)

1500 V

250 V

250 V

Max. pulse energy (10 μs)

30 mWs

15 mWs

15 mWs

1 dB compression of input mixer (RF attenuation 0 dB)

Analyzer mode

+10 dBm nominal

Intermodulation

3rd-order intercept point (TOI)

Analyzer mode,

Δ

f > 5 × IF bandwidth or resolution bandwidth,

or >10 kHz

12 dBm, typ. 15 dBm for f > 150 MHz

12 dBm, typ. 15 dBm for

f > 150 MHz;

10 dBm for

f > 7 GHz

Receiver mode, preamplifi er off

2 dBm, typ. 5 dBm for f > 150 MHz

Receiver mode, preamplifi er on

–18 dBm, typ. –15 dBm for f > 150 MHz

Intercept point k2, analyzer mode

>25 dBm, typ. for f < 150 MHz
>40 dBm, typ. for f > 150 MHz

Level display (receiver mode)

Digital

numeric, 0.1 dB resolution

Analog

bargraph display, separate for each detector, max. 4 simultaneously

Spectrum

level axis 10 dB to 200 dB in steps of 10 dB, frequency axis user-selectable, linear or
logarithmic

Units of level display

dBμV, dBm, dBμA, dBpW, dBpT, dB(μV/m), dB(μA/m), dB

x

4)

/MHz

Detectors

average (AV), RMS, peak (PK), quasi-peak (QP) and CISPR-average (CISPR AV); 4 detectors can
be switched on simultaneously

Measurement time

100 μs to 100 s, selectable

Level display (analyzer mode)

Result display

500 × 400 pixels (with one diagram displayed), max. 2 diagrams with independent settings

Logarithmic level range

10 dB to 200 dB in steps of 10 dB

Linear level range

10 % of reference level per division (10 divisions) or logarithmic scaling

Traces

max. 4 traces with one diagram (2 traces per diagram with 2 diagrams); quasi-analog display
of all traces

Trace detectors

max. peak, min. peak, auto peak (normal), sample, RMS, average

Trace functions

clear/write, max. hold, min. hold, average

Setting range of reference level

Logarithmic level display

–130 dBm to 30 dBm in steps of 0.1 dB

Linear level display

7.0 nV to 7.07 V in steps of 1 %

Unit of level axis

dBm, dBμV, dBμA, dBpW, dB

x

4)

/MHz (logarithmic level display);

mV, μA, pW, nW (linear level display)

ESIB_dat_en.indd 10