beautypg.com

Specifi cations – Atec Rohde-Schwarz-ESIB Series User Manual

Page 8

background image

8

EMI Test Receivers ¸ESIB

Specifi cations

Specifi cations apply under the following conditions: 30 minutes warm-up time at ambient temperature, specifi ed environmental conditions met, calibration cycle
adhered to, and total calibration performed. Data without tolerances: typical values only. Data designated “nominal” applies to design parameters and is not tested.

¸ESIB 7

¸ESIB 26

¸ESIB 40

Frequency

Frequency range

Input 1
Input 2

20 Hz to 7 GHz
20 Hz to 1 GHz

20 Hz to 26.5 GHz
20 Hz to 1 GHz

20 Hz to 40 GHz
20 Hz to 1 GHz

Frequency resolution

0.01 Hz

Internal reference frequency (nominal)

Aging per day

1)

1 × 10

–9

Aging per year

1)

2 × 10

–7

Temperature drift (0 °C to 50 °C)

5 × 10

–8

Total error (per year)

2.5 × 10

–7

External reference frequency

10 MHz or n × 1 MHz, n = 1 to 16

Frequency display (receiver mode)

Display

numeric display

Resolution

0.1 Hz

Frequency display (analyzer mode)

Display

with marker

Resolution
Accuracy (sweep time >3 × auto sweep time)

0.1 Hz to 10 kHz (depending on span)
± (marker frequency × reference error + 0.5 % × span + 10 % × resolution bandwidth +
½ (last digit))

Frequency counter

measures the marker frequency

Resolution

0.1 Hz to 10 kHz, selectable

Count accuracy (S/N >25 dB)

± (frequency × reference error + ½ (last digit))

Display range for frequency axis

0 Hz, 10 Hz to 7 GHz

0 Hz, 10 Hz to 27 GHz

0 Hz, 10 Hz to 40 GHz

Resolution/accuracy of display range

0.1 Hz/±1 %

Spectral purity

for frequencies >500 MHz: see diagram on page 7

SSB phase noise, f

500 MHz

Carrier offset

100 Hz
1 kHz
10 kHz
100 kHz

2)

1 MHz

2)

<–81 dBc (1 Hz)
<–100 dBc (1 Hz)
<–114 dBc (1 Hz)
<–111 dBc (1 Hz)
<–129 dBc (1 Hz)

Frequency scan (receiver mode)

Scan

scan with max. 10 subranges with different settings

Measurement time per frequency

100 μs to 100 s, selectable

Sweep (analyzer mode)

Span 0 Hz (zero span)

1 μs to 2500 s, selectable in steps of 5 %

Span

10 Hz

5 ms to 1000 s, selectable in steps of

10 %

Accuracy

±1 %

Picture refresh rate
(span

7 GHz)

>20 updates/s with 1 trace
>15 updates/s with 2 traces at shortest sweep time

Sampling rate

50 ns (20 MHz A/D converter)

Number of pixels

500

Time measurement

with marker and cursor lines

Resolution

50 ns

ESIB_dat_en.indd 8

04.01.2006 13:27:23