Atec Rohde-Schwarz-ESCI Series User Manual
Page 6
Version 03.00, June 2009
6
Rohde & Schwarz
R&S
®
ESCI/ESCI7 EMI Test Receiver
Noise indication (receiver mode)
Nominal, calculated from DANL data, 0 dB RF attenuation, 50 Ω termination
without preamplifier
R&S
®
ESCI, R&S
®
ESCI7
9 kHz, BW = 200 Hz
< 5 dBμV
150 kHz, BW = 200 Hz
< 0 dBμV
150 kHz, BW = 9 kHz
< 16 dBμV
1 MHz, BW = 9 kHz
< –4 dBμV
10 MHz to 30 MHz, BW = 9 kHz
< –6 dBμV
30 MHz to 1 GHz, BW = 120 kHz
< 6 dBμV
1 GHz to 3 GHz, BW = 1 MHz
< 16 dBμV
R&S
®
ESCI7
3 GHz to 7 GHz, BW = 1 MHz
< 20 dBμV
with preamplifier
R&S
®
ESCI, R&S
®
ESCI7
9 kHz, BW = 200 Hz
< –15 dBμV
150 kHz, BW = 200 Hz
< –20 dBμV
150 kHz, BW = 9 kHz
< –4 dBμV
1 MHz, BW = 9 kHz
< –14 dBμV
10 MHz to 30 MHz, BW = 9 kHz
< –16 dBμV
30 MHz to 1 GHz, BW = 120 kHz
< –4 dBμV
1 GHz to 3 GHz, BW = 1 MHz
< 6 dBμV
R&S
®
ESCI7
Average (AV) display
3 GHz to 7 GHz, BW = 1 MHz
< 3 dBμV
max peak
typ. +11 dB
RMS
typ. +1 dB
quasi-peak
band A
typ. +3 dB
band B
typ. +4 dB
Increase of DANL relative to AV display
bands C and D
typ. +6 dB
Immunity to interference
Image frequency
> 70 dB
Intermediate frequency
> 70 dB
Spurious response
f > 1 MHz, 0 dB RF attenuation,
without input signal
< –103 dBm
Other interfering signals
∆f > 100 kHz, mixer level < –10 dBm
< –70 dBc
RF shielding
field strength 3 V/m, 0 dB RF attenuation,
50 Ω termination, f ≠ f
IF
level indication < 10 dBμV, nominal
Level display (receiver mode)
digital numeric,
resolution 0.01
dB
Level display
analog
bargraph display separate for each
detector
level axis
10 dB to 200 dB in steps of 10 dB
Spectrum
frequency axis
linear or logarithmic selectable
Detectors
Three detectors can be switched on
simultaneously.
average (AV), RMS, max peak, min peak,
quasi-peak (QPK), CISPR-AV, CISPR-RMS
Units of level display
dBμV, dBm, dBμA, dBpW, dBpT
Measurement time
selectable
33 μs to 100 s
Level display (analyzer mode)
Screen
501 × 400 pixels
(one measurement diagram);
max. two measurement diagrams
with independent settings
Logarithmic level display range
1 dB, 10 dB to 200 dB in steps of 10 dB
Linear level display range
10 % of reference level per level division,
10 divisions
one measurement diagram
3
Number of traces
two measurement diagrams
6
Trace detectors
max peak, min peak, auto peak, sample,
quasi-peak, average, RMS
Trace functions
clear/write, max hold, min hold, average
default value
501
Number of measurement points
range
125 to 8001 in steps of approx. a factor of 2