beautypg.com

Scan (receiver mode), Sweep (analyzer mode), Resolution bandwidths – Atec Rohde-Schwarz-ESCI Series User Manual

Page 3: Preselection

background image

Version 03.00, June 2009

Rohde & Schwarz

R&S

®

ESCI/ESCI7 EMI Test Receiver 3

Scan (receiver mode)

Scan

scan of max. 10 subranges with different,
independent settings

Measurement time per frequency

selectable

33 μs to 100 s

Sweep (analyzer mode)

in time domain, span = 0 Hz

1 μs to 16000 s

resolution 125

ns

Sweep time

in frequency domain, span ≥ 10 Hz

2.5 ms to 16000 s

Max. deviation of sweep time

1 %

Resolution bandwidths

Sweep filters
3 dB bandwidths

10 Hz to 3 MHz, in steps of 1/3/10

≤ 100 kHz

< 3 %

Bandwidth accuracy

300 kHz to 3 MHz

< 10 %

≤ 100 kHz

< 5

Shape factor 60 dB:3 dB

300 kHz to 3 MHz

< 15

6 dB bandwidths

200 Hz, 9 kHz, 120 kHz

EMI bandwidths

pulse bandwidth

1 MHz

≤ 120 kHz

< 3 %

Bandwidth accuracy

1 MHz

< 10 %

≤ 120 kHz

< 5

Shape factor 60 dB:6 dB

1 MHz

< 15

Video bandwidths

analyzer mode

1 Hz to 10 MHz, in steps of 1/3/10

FFT filters

analyzer mode

3 dB bandwidths

1 Hz to 30 kHz, in steps of 1/3/10

Bandwidth accuracy

5 %, nominal

Shape factor 60 dB:3 dB

2.5, nominal

Channel filters
Bandwidths

100/200/300/500

Hz;

1/1.5/2/2.4/2.7/3/3.4/4/4.5/5/6/8.5/9/10/
12.5/14/15/16/18 (RRC)/20/21/24.3 (RRC)/
25/30/50/100/150/192/200/300/500 kHz
1/1.228/1.28 (RRC)/1.5/2/3/3.84 (RRC)/
4.096 (RRC)/ 5 MHz
(RRC = root raised cosine)

Preselection

Preselection

can be switched off in analyzer mode

R&S

®

ESCI: 11 preselection filters

R&S

®

ESCI7: 12 preselection filters

Bandwidths (–6 dB), nominal

R&S

®

ESCI, R&S

®

ESCI7

< 150 kHz

230 kHz, fixed-tuned lowpass filter

150 kHz to 2 MHz

2.6 MHz, fixed-tuned bandpass filter

2 MHz to 8 MHz

2 MHz, tracking bandpass filter

8 MHz to 30 MHz

6 MHz, tracking bandpass filter

30 MHz to 70 MHz

15 MHz, tracking bandpass filter

70 MHz to 150 MHz

30 MHz, tracking bandpass filter

150 MHz to 300 MHz

60 MHz, tracking bandpass filter

300 MHz to 600 MHz

80 MHz, tracking bandpass filter

600 MHz to 1 GHz

100 MHz, tracking bandpass filter

1 GHz to 2 GHz

tracking highpass filter

2 GHz to 3 GHz

fixed-tuned highpass filter

R&S

®

ESCI7

3 GHz to 7 GHz

tracking bandpass filter

Preamplifier

switchable, between preselection and
1st mixer

20 dB