Atec Rohde-Schwarz-ESCI Series User Manual
Page 5
Version 03.00, June 2009
Rohde & Schwarz
R&S
®
ESCI/ESCI7 EMI Test Receiver 5
RF attenuation 0 dB, RBW = 10 Hz,
VBW = 1 Hz, span = 0 Hz, trace average function over 20 sweeps, 50 Ω termination
without preselection, without preamplifier, AC-coupled
R&S
®
ESCI
9 kHz
< –105 dBm, nominal
100 kHz
< –110 dBm, nominal
1 MHz
< –130 dBm, nominal
10 MHz to 1 GHz
< –142 dBm, typ. –145 dBm
1 GHz to 2.5 GHz
< –140 dBm, typ. –143 dBm
2.5 GHz to 3 GHz
< –138 dBm, typ. –141 dBm
R&S
®
ESCI7
1 MHz
< –128 dBm, nominal
10 MHz to 1 GHz
< –140 dBm, typ. –143 dBm
1 GHz to 2.5 GHz
< –138 dBm, typ. –141 dBm
2.5 GHz to 3 GHz
< –136 dBm, typ. –139 dBm
3 GHz to 7 GHz
< –138 dBm, typ. –141 dBm
without preselection, without preamplifier, DC-coupled
R&S
®
ESCI
9 kHz
< –115 dBm
100 kHz
< –120 dBm
1 MHz
< –140 dBm, typ. –143 dBm
10 MHz to 1 GHz
< –142 dBm, typ. –145 dBm
1 GHz to 2.5 GHz
< –140 dBm, typ. –143 dBm
2.5 GHz to 3 GHz
< –138 dBm, typ. –141 dBm
R&S
®
ESCI7
9 kHz
< –115 dBm
100 kHz
< –120 dBm
1 MHz
< –138 dBm, typ. –141 dBm
10 MHz to 1 GHz
< –140 dBm, typ. –143 dBm
1 GHz to 2.5 GHz
< –138 dBm, typ. –141 dBm
2.5 GHz to 3 GHz
< –136 dBm, typ. –139 dBm
3 GHz to 7 GHz
< –138 dBm, typ. –141 dBm
with preselection, without preamplifier, DC-coupled
R&S
®
ESCI
9 kHz
< –115 dBm
100 kHz
< –120 dBm, typ. –140 dBm
1 MHz
< –140 dBm, typ. –148 dBm
10 MHz to 1 GHz
< –142 dBm, typ. –150 dBm
1 GHz to 2.5 GHz
< –140 dBm, typ. –148 dBm
2.5 GHz to 3 GHz
< –138 dBm, typ. –141 dBm
R&S
®
ESCI7
9 kHz
< –115 dBm
100 kHz
< –120 dBm, typ. –140 dBm
1 MHz
< –138 dBm, typ. –146 dBm
10 MHz to 1 GHz
< –140 dBm, typ. –148 dBm
1 GHz to 2.5 GHz
< –138 dBm, typ. –146 dBm
2.5 GHz to 3 GHz
< –136 dBm, typ. –139 dBm
3 GHz to 7 GHz
< –138 dBm, typ. –141 dBm
with preselection, with preamplifier, DC-coupled
R&S
®
ESCI
9 kHz
< –135 dBm
100 kHz
< –140 dBm
1 MHz
< –150 dBm, typ. –153 dBm
10 MHz to 1 GHz
< –152 dBm, typ. –155 dBm
Displayed average noise level (DANL)
(analyzer mode)
1 GHz to 3 GHz
< –150 dBm, typ. –153 dBm
R&S
®
ESCI7
9 kHz
< –135 dBm
100 kHz
< –140 dBm
1 MHz
< –148 dBm, typ. –151 dBm
10 MHz to 1 GHz
< –150 dBm, typ. –153 dBm
1 GHz to 7 GHz
< –148 dBm, typ. –151 dBm