beautypg.com

Atec Rohde-Schwarz-ESCI Series User Manual

Page 5

background image

Version 03.00, June 2009

Rohde & Schwarz

R&S

®

ESCI/ESCI7 EMI Test Receiver 5

RF attenuation 0 dB, RBW = 10 Hz,
VBW = 1 Hz, span = 0 Hz, trace average function over 20 sweeps, 50 Ω termination
without preselection, without preamplifier, AC-coupled

R&S

®

ESCI

9 kHz

< –105 dBm, nominal

100 kHz

< –110 dBm, nominal

1 MHz

< –130 dBm, nominal

10 MHz to 1 GHz

< –142 dBm, typ. –145 dBm

1 GHz to 2.5 GHz

< –140 dBm, typ. –143 dBm

2.5 GHz to 3 GHz

< –138 dBm, typ. –141 dBm

R&S

®

ESCI7

1 MHz

< –128 dBm, nominal

10 MHz to 1 GHz

< –140 dBm, typ. –143 dBm

1 GHz to 2.5 GHz

< –138 dBm, typ. –141 dBm

2.5 GHz to 3 GHz

< –136 dBm, typ. –139 dBm

3 GHz to 7 GHz

< –138 dBm, typ. –141 dBm

without preselection, without preamplifier, DC-coupled

R&S

®

ESCI

9 kHz

< –115 dBm

100 kHz

< –120 dBm

1 MHz

< –140 dBm, typ. –143 dBm

10 MHz to 1 GHz

< –142 dBm, typ. –145 dBm

1 GHz to 2.5 GHz

< –140 dBm, typ. –143 dBm

2.5 GHz to 3 GHz

< –138 dBm, typ. –141 dBm

R&S

®

ESCI7

9 kHz

< –115 dBm

100 kHz

< –120 dBm

1 MHz

< –138 dBm, typ. –141 dBm

10 MHz to 1 GHz

< –140 dBm, typ. –143 dBm

1 GHz to 2.5 GHz

< –138 dBm, typ. –141 dBm

2.5 GHz to 3 GHz

< –136 dBm, typ. –139 dBm

3 GHz to 7 GHz

< –138 dBm, typ. –141 dBm

with preselection, without preamplifier, DC-coupled

R&S

®

ESCI

9 kHz

< –115 dBm

100 kHz

< –120 dBm, typ. –140 dBm

1 MHz

< –140 dBm, typ. –148 dBm

10 MHz to 1 GHz

< –142 dBm, typ. –150 dBm

1 GHz to 2.5 GHz

< –140 dBm, typ. –148 dBm

2.5 GHz to 3 GHz

< –138 dBm, typ. –141 dBm

R&S

®

ESCI7

9 kHz

< –115 dBm

100 kHz

< –120 dBm, typ. –140 dBm

1 MHz

< –138 dBm, typ. –146 dBm

10 MHz to 1 GHz

< –140 dBm, typ. –148 dBm

1 GHz to 2.5 GHz

< –138 dBm, typ. –146 dBm

2.5 GHz to 3 GHz

< –136 dBm, typ. –139 dBm

3 GHz to 7 GHz

< –138 dBm, typ. –141 dBm

with preselection, with preamplifier, DC-coupled

R&S

®

ESCI

9 kHz

< –135 dBm

100 kHz

< –140 dBm

1 MHz

< –150 dBm, typ. –153 dBm

10 MHz to 1 GHz

< –152 dBm, typ. –155 dBm

Displayed average noise level (DANL)
(analyzer mode)

1 GHz to 3 GHz

< –150 dBm, typ. –153 dBm

R&S

®

ESCI7

9 kHz

< –135 dBm

100 kHz

< –140 dBm

1 MHz

< –148 dBm, typ. –151 dBm

10 MHz to 1 GHz

< –150 dBm, typ. –153 dBm

1 GHz to 7 GHz

< –148 dBm, typ. –151 dBm