beautypg.com

General data – Atec Rohde-Schwarz-ESCI Series User Manual

Page 10

background image

Version 03.00, June 2009

10

Rohde & Schwarz

R&S

®

ESCI/ESCI7 EMI Test Receiver

General data

Display

21 cm TFT color display

Resolution

640 × 480 pixel (VGA)

Pixel error rate

< 2 × 10

–5

Mass memory

1.44 Mbyte 3½'' disk drive, hard disk

Data storage

R&S

®

ESCI only

> 500 instrument setups and traces

Temperature ranges

+5 °C to + 40 °C

Operating temperature range

with R&S

®

ESCI-B20 option

0 °C to +50 °C

+5 °C to + 45 °C

Permissible temperature range

with R&S

®

ESCI-B20 option

0 °C to +55 °C

Storage temperature range

–40 °C to +70 °C

Climatic loading

+40 °C at 95 % relative humidity
(EN 60068-2-30)

Mechanical resistance
Sinusoidal vibration

0.5 g from 5 Hz to 150 Hz,
max. 2 g at 55 Hz,
in line with EN 60068-2-6, EN 61010-1,
MIL-T-28800D, class 5

10 Hz to 100 Hz, acceleration 1 g (rms)

Random vibration

with R&S

®

ESCI-B20 option

10 Hz to 300 Hz, acceleration 1.9 g (rms)

Shock

40 g shock spectrum,
in line with MIL-STD-810C and
MIL-T-28800D, classes 3 and 5

operation with external reference

2 years

Recommended calibration interval

operation with internal reference

1 year

Power supply
AC supply

100 V to 240 V AC, 50 Hz to 400 Hz,
3.1 A to 1.3 A,
class of protection I in line with VDE 411

Power consumption

typ. 70 VA

Safety

in line with EN 61010-1, UL 3111-1,
CSA C22.2 No. 1010-1, IEC 1010-1

EMC

EMC Directive 2004/108/EC
including:

EN 61326 class B (emission),
CISPR 11/EN 55011 group 1 class B
(emission)
EN 61326 table A.1 (immunity,
industrial)

Test marks

VDE, GS, CSA, CSA-NRTL/C

Dimensions and weight
Dimensions

W × H × D

412 mm × 197 mm × 417 mm
(16.22 in × 7.76 in × 16.42 in)

R&S

®

ESCI

10.5 kg (23.15 lb)

Weight without options

R&S

®

ESCI7

12.4 kg (27.34 lb)